12-128 Instrument Control Library
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
status.system3.*
status.system3.condition
status.system3.enable
status.system3.event
* = condition, enable or event
Attribute
TSP-Link system3 data structure register set.
Usage
Reads condition, enable and event registers:
enablereg = status.system3.condition
enablereg = status.system3.enable
enablereg = status.system3.event
Writes to enable register:
status.system3.enable = enablereg
Set
enablereg
to one of the following values:
0
Clears all bits.
1
or
status.system3.EXTENSION_BIT
Sets EXT bit (B0).
1
or
status.system3.EXT
Sets EXT bit (B0).
status.system3.NODEn
Sets a node bit (Bn);
n
= 29 to 42.
Remarks
• In an expanded system (TSP-Link), this attribute is used to read or write to the
system3 node registers.
• Reading a system3 node register returns a numeric value whose binary equivalent
indicates which register bits are set. The bits of the system3 node register are
identified as follows:
Bit B0 – EXT bit
Bit B4 – Node 32
Bit B8 – Node 36
Bit B12 – Node 40
Bit B1 – Node 29
Bit B5 – Node 33
Bit B9 – Node 37
Bit B13 – Node 41
Bit B2 – Node 30
Bit B6 – Node 34
Bit B10 – Node 38
Bit B14 – Node 42
Bit B3 – Node 31
Bit B7 – Node 35
Bit B11 – Node 39
Bit B15 – Not used
For example, assume value 9 is returned for the enable register. The binary
equivalent is 0000000000001001. This value indicates that bit 0 (EXT) and bit 3
(Node 31) are set.
• Assigning a value to the
status.system3.enable
attribute sets the extension bit or
a node bit of the system3 node enable register.
Details
See “
System Summary Event Registers
Also see
Example
Sets the extension bit of the system3 enable register:
status.system3.enable = status.system3.EXT
Summary of Contents for SourceMeter 2600 Series
Page 575: ...AppendixF DisplayCharacterCodes Appendix F topics Introduction page F 2...
Page 592: ......
Page 593: ......