Series 2600 System SourceMeters Reference Manual
Factory Scripts 13-5
Return to
2600S-901-01 Rev. A / May 2006
TSP Project Name:
KIFactoryGeneral
TSP Test Script Name:
KIGeneral
Firmware Version:
1.0.2 and later
Function:
SweepILinMeasureV(smu, starti, stopi, stime, points)
Description: Performs a linear current sweep with voltage measured at every step (point):
• Sets the
smu
to output
starti
amps, allows the source to settle for
stime
seconds and then performs a voltage measurement.
• Sets the
smu
to output the next amps step, allows the source to settle for
stime
seconds and then performs a voltage measurement.
• Repeats the above sequence until the voltage is measured on the
stopi
amps step.
The linear step size is automatically calculated as follows:
step = (
stopi
–
starti
) / (
points
– 1)
starti
= Voltage measurement
stopi
stime
stime
stime
Parameters:
smu, starti, stopi, stime, points
smu:
SourceMeter Channel (A or B). Defaults to smua if all parameters
are omitted when function is called.
starti:
Sweep start current in amps.
stopi:
Sweep stop current in amps.
stime:
Settling time in seconds. Occurs after stepping the source and
before performing a measurement.
points:
Number of sweep points (must be
≥
2).
Data: Voltage measurements, current source values and timestamps are stored in
smuX.nvbuffer1
.
Example:
SweepILinMeasureV(smua, -1E-3, 1E-3, 0, 100)
This function performs a 100-point linear current sweep starting at -1mA and
stopping at +1mA. Voltage is measured at every step (point) in the sweep.
Since
stime
is set for 0s, voltage will be measured as fast as possible after
each current step.