2600S-901-01 Rev. A / May 2006
Flash firmware upgrade
As Keithley Instruments develops additional factory scripts, they will be made
available on the Keithley Instruments web site (www.keithley.com) as a flash
firmware upgrade for the Series 2600.
CAUTION
External circuitry connected to input/output terminals while
attempting a flash upgrade may cause instrument and/or
DUT damage. Disconnect input/output terminals before
performing a flash upgrade.
After downloading the new flash file from the Keithley Instruments website, use
the Test Script Builder (TSB) to upgrade the firmware of your Series 2600:
1.
On the PC desktop, double-click the icon for the Test Script Builder.
2. On the Instrument Console toolbar, click the Open Instrument icon and then
select your communication interface from the Select Instrument Resource
dialog box. Details on opening communications are provided in
.
Open Instrument
icon
3. On the Instrument Console toolbar, click the Menu icon to display the menu.
Menu
icon
4. From the drop-down menu, select Instrument and then click Flash.
5. From the Select A Firmware Data File dialog box, use the browser to select
the File name of the new firmware and click Open to upgrade the firmware of
the Series 2600.