14-4 Display Operations
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
Display resolution
Display resolution for measured readings can be set to 4-1/2, 5-1/2 or 6-1/2 digit
resolution.
The
display.smuXdigits
attribute is used to set display resolution for measured
readings:
display.smuX.digits = digits
Where:
smuX
=
smua
or
smub
digits
is set to one of the following values:
4
or
display.DIGITS_4_5
5
or
display.DIGITS_5_5
6
or
display.DIGITS_6_5
Display resolution example:
The following command sets SMU A for 5-1/2 digit resolution for measured
readings:
display.smua.digits = display.DIGITS_5_5
Display messages
The display of the Series 2600 can be used to display user-defined messages.
For example, while a test is running, the following message can be displayed on
the Series 2600.
Test in Process
Do Not Disturb
The top line of the display can accommodate up to 20 characters (including
spaces). The bottom line can display up to 32 characters (including spaces) at a
time.
NOTE
Most of the display functions and attributes that are
associated with display messaging will automatically
select the user screen. The attribute for the display
screen is explained in “
” in this
section.
The reset functions (
reset
or
smuX.reset
) have no effect
on the defined display message or its configuration, but
will set the display mode back to the previous source-
measure display mode.