Series 2600 System SourceMeters Reference Manual
Common Commands
C-3
Return to
2600S-901-01 Rev. A / May 2006
Script command equivalents
Script command equivalents for the common commands in
are
summarized in
. See Section 12 for details on script commands.
Table C-2
Script command equivalents
Common
command
Script command equivalent
*CLS
status.reset()
*ESE?
print(tostring(status.standard.enable))
*ESE <mask>
status.standard.enable = <mask>
*ESR?
print(tostring(status.standard.event))
*IDN?
print([[Keithley Instruments Inc., Model]]..localn-
ode.model..
[[, ]]..localnode.serialno..[[, ]]..localnode.revision)
*OPC?
waitcomplete() print([[1]])
*OPC
opc()
*RST
reset()
*SRE?
print(tostring(status.request_enable))
*SRE <mask>
status.request_enable = <mask>
*STB?
print(tostring(status.condition))
*TRG
N/A
*TST?
print([[0]])
*WAI
waitcomplete()