Series 2600 System SourceMeters Reference Manual TSP Programming Fundamentals 2-29
Return to
2600S-901-01 Rev. A / May 2006
Instrument Console
With communications established with the SourceMeter, the Instrument Console
is used for the following operations:
• Execute chunks, which are individual ICL commands and TSL
programming statements.
• Display returned data (readings and messages).
• Display error messages caused by erroneous code sent from the
Instrument Console.
The instrument console is opened by clicking the Instrument Console tab in the
lower window pane of the Test Script Builder (see
An active Instrument Console displays the TSP> prompt. Type in a command
after the prompt and press Enter to execute it. For example, type in the following
command:
TSP>reset ()
After pressing Enter, the SourceMeter will reset to its default settings.
Code and messages in the Instrument Console can be cleared by clicking the
Clear Console Window icon shown in . It can also be cleared from the mouse
menu as follows: Position the mouse pointer in the console window, right-click the
mouse and then select Clear Console Window from the mouse menu.
Instrument Console icons
After communications with the SourceMeter are open, all of the icons on the
Instrument Console toolbar (shown in ) will be active.
Figure 2-18
Instrument Console icons
NOTE See “
Sending commands and statements
” for example code
that can be sent from the Instrument Console.
Close Instrument
Clear Console Window
Abort Execution
Reset
Send Software Trigger
Delete a Script From NVRAM
Menu