Table 4-2
Maximum compliance values
Model 2601/2602
Model 2611/2612
Model 2635/2636
Source
range
Maximum
compliance
value
Source
range
Maximum
compliance
value
Source
range
Maximum
compliance
value
100mV
1V
6V
40V
3A
3A
3A
1A
200mV
2V
20V
200V
1.5A
1.5A
1.5A
100mA
200mV
2V
20V
200V
1.5A
1.5A
1.5A
100mA
100nA
1µA
10µA
100µA
1mA
10mA
100mA
1A
3A
40V
40V
40V
40V
40V
40V
40V
40V
6V
100nA
1µA
10µA
100µA
1mA
10mA
100mA
1A
1.5A
200V
200V
200V
200V
200V
200V
200V
20V
20V
1nA
10nA
100nA
1µA
10µA
100µA
1mA
10mA
100mA
1A
1.5A
200V
200V
200V
200V
200V
200V
200V
200V
200V
20V
20V
4-4
Return to
2600S-901-01 Rev. C / January 2008
Section 4: Basic Operation
Series 2600 System SourceMeter® Instruments Reference Manual
*smuX = smua for the Model 2601/2611/2635; smuX = smua (Channel A) or smub (Channel B) for
the Model 2602/2612/2636.
Setting the compliance limit
Front panel compliance limit
Set the compliance limit from the front panel as follows:
1. For the Model 2601/2611/2635 or the Model 2602/2612/2636 single-channel display mode,
press the LIMIT key to directly access compliance editing.
2. For the Model 2602/2612/2636 dual-channel display mode, press CONFIG then LIMIT, then
select CURRENT or VOLTAGE as desired. Press ENTER or the Rotary Knob.
3. Press the Rotary Knob, set the compliance limit to the desired value, and then press
ENTER or the Rotary Knob to complete editing.
4. Press EXIT to return to the normal display.