2600S-901-01 Rev. C / January 2008
Return to
8-9
Series 2600 System SourceMeter® Instruments Reference Manual
Section 8: Source-Measure Concepts
Figure 8-4
Model 2611/2612/2635/2636 continuous power operating boundaries
-V
+V
-200V
(IV)
Sink
+I
1.5A
100mA
-20V
20V
200V
(III)
Source
(II)
Sink
-100mA
-1.5A
-I
(I)
Source
Energy storage DUT
Energy storage DUT
£
30°C
T
amb
I-Source operating boundaries
Model 2601/2602 I-Source operating boundaries
shows the operating boundaries for the I-Source. Only the first quadrant of operation is
covered. Operation in the other three quadrants is similar.
A shows the output characteristics for the I-Source. As shown, the Model 2601/2602
SourceMeter can output up to 1.01A at 40V, or 3.03A at 6V. Note that when sourcing more than
1.01A, voltage is limited to 6V.
B shows the limit lines for the I-Source. The current source limit line represents the
maximum source value possible for the presently selected current source range. The voltage
compliance limit line represents the actual compliance that is in effect. See
earlier in this section. These limit lines are boundaries that represent the operating limits of the
SourceMeter for this quadrant of operation. The operating point can be anywhere inside (or on)
these limit lines. The limit line boundaries for the other quadrants are similar.