Editing controls
Source and compliance editing
When the Series 2600 is in the edit mode (EDIT annunciator on), the editing controls are used to
set source and compliance values. Note that source auto ranging will turn off when editing the
source value.
Editing source values
Complete the following steps to edit the source.
1. Press the
SRC
key.
The cursor flashes in the source value field.
2. Use the
CURSOR
arrow keys to move the cursor to the desired digit using.
3. Push the
Wheel
in to edit the source value.
The Edit indicator displays.
4. Do one of the following to change the source value:
• Rotate the wheel to adjust the digit.
Note: The digit is automatically overflow or underflow to the next digit if the maximum
source value is reached.
• Use the numeric keys (0-9, +/-, 0000) to enter the source value.
Note
: The +/- toggles the polarity, and 0000 sets the value to 0.
5. Once the desired value displays, press
ENTER
.
Note: The Edit indicator does not display.
6. (Optional) Press the
EXIT
key to cancel source editing.
Editing compliance values
Complete the following steps to edit the compliance value.
1. Do one of the following:
• (Model 2601/2611/2635 and 2602/2612/2636 in single-channel display mode only) Press
the
LIMIT
key.
• (Model 2602/2612/2636 dual-channel display mode only) Press
LIMIT
or
CONFIG,
then
LIMIT
to edit the compliance limit.
2. Choose one of the following:
•
VOLTAGE
•
CURRENT
3. Use the
CURSOR
arrow keys to move the cursor to the desired value.
4. Press in on the wheel to enter edit mode.
The Edit indicator displays.
5. Do one of the following to modify the compliance limit value:
• Rotate the wheel to adjust the value.
Note
: The value is automatically overflow or underflow to the next value if the minimum or
maximum value is reached.
• Use the numeric keys (0-9) to enter the value.
6. Press
ENTER
to complete editing.
Note
: The Edit indicator does not display.
7. (Optional) Press the
EXIT
key cancel changes.