Table 16-1
Recommended calibration equipment
Description
Manufacturer/Model
Accuracy
1G
Ω
, 200V, 1% stan-
dard
Keithley Instruments Model
2600-STD-RES
Resistance
6
1G
Ω
:
XXXPPM
Digital Multimeter
Agilent 3458A
DC Voltage
1
(2601/2602)
DC Voltage
2
(2611/2612)
DC current
3
90mV:
0.9V:
5.4V:
36V:
190mV:
1.8V:
18V:
180V:
90nA:
0.9µA:
9µA:
90µA:
0.9mA:
9mA:
90mA:
0.9A:
±8ppm
±5ppm
±4ppm
±6ppm
±5ppm
±4ppm
±6ppm
±6ppm
±430ppm
±45ppm
±25ppm
±23ppm
±20ppm
±20ppm
±35ppm
±110ppm
0.5
Ω
, 250W, 0.1%
Precision Resistor
Isotek
RUG-Z-R500-0.1-TK3
Resistance
4
0.5
Ω
:
±125ppm
50
Ω
Resistors (2)
Any suitable.
5
1G
Ω
, 200V, 1%
standard
Keithley
2600-STD-RES
Resistance
6
1G
Ω
:
+/-250ppm
1. 90-day specifications show full-range accuracy of recommended model used for specified
calibration point.
2. Id.
3. Id.
4. Resistor used to calibrate Model 2601/2602 3A and 10A ranges and Model 2611/2612/2635/
2636 1.5A and 10A ranges should be characterized to uncertainty shown using resistance
function of a digital multimeter before use.
5. Used for contact check calibration. Characterize resistors using ohms function of digital
multimeter before use.
6. Standard is a guarded and characterized 1G
Ω
resistor used to test Model 2635/2636 100pA to
100nA current ranges.
16-4
Return to
2600S-901-01 Rev. C / January 2008
Section 16: Calibration
Series 2600 System SourceMeter® Instruments Reference Manual