2600S-900-01 Rev. C / January 2008
Return to
1-17
Series 2600 System SourceMeter
Section 1: Front and Rear Panel Operation
Step 4: Turn on the output
Press the
CHAN A
(Model 2602/2612/2636)
OUTPUT ON/OFF
key to turn on the source output.
Step 5: Store readings
1. Press the
STORE
key to enable storage.
The * (asterisk) indicator displays.
2. Press
EXIT
to stop data storage.
Step 6: Turn off the output
• Press the
OUTPUT ON/OFF
key to turn off the output after storage has completed.
Step 7: Recall readings
1. Press the
RECALL
key to access buffer readings. (Repeatedly pressing
RECALL
will cycle
through Buffer 1 then Buffer 2 for Channel A and then Channel B). A message will be
displayed if a buffer is empty.
2. Note the buffer display data (see
):
• The reading is on the top display at the left.
• The buffer location number is on the right. For example, location #000001 indicates that
the displayed reading is stored at the first memory location.
• The source value is positioned at the lower left side of the display.
• The timestamp is positioned at the lower right side. The first source-measure reading
stored in the buffer (#000001) is time stamped at 00000000.001 seconds. Subsequent
readings are time stamped relative to when the time storage was started. The interval
between readings will depend on the reading rate.
3. To display the other readings stored in the buffer, choose the desired memory location
number:
• Use the to increment and decrement the selected digit of the location number.
• Set the cursor position with the or
CURSOR
keys.
4. To exit from the data store recall mode, press
EXIT
.
Figure 1-7
Buffer display format
1.00000 mA #0000001
SrcA1:+10.0000 V @00000000.001s
Use Knob
Use Knob
Use Knob
Buffer Location Number
Timestamp (seconds)
Reading
Source Value
1.00000 mA #0000002
SrcA1:+10.0000 V @00000000.002s
1.00000 mA #0000003
SrcA1:+10.0000 V @00000000.003s
1.00000 mA #0000004
SrcA1:+10.0000 V @00000000.004s
Note: Source values and timestamps
will be displayed only if enabled when
buffer is configured.
(SrcA1=