2600S-900-01 Rev. C / January 2008
Return to
2-3
Series 2600 System SourceMeter
Section 2: Remote Operation
Step 3: Configure the interface
GPIB interface configuration
Set the primary address using the following procedure. The Series 2600 primary address must be
the same as that specified in your program, or the two devices will not be able to communicate.
1. Press
MENU
>
GPIB
and then press the
ENTER
.
2. Select
ADDRESS
and then press the
ENTER
.
The Address screen displays.
3. Turn the
navigation wheel to the right or to the left to select the desired number.
4. Push in on the
navigation wheel
and then turn the
navigation wheel
to the left or to the
right to choose an address.
NOTE The address can range from 1-30.
5. Press the
navigation wheel
to set the address.
6. Press
EXIT
to until the Main Menu displays.
RS-232 interface configuration
Set RS-232 parameters as covered below. The Series 2600 RS-232 parameters must agree with
those of the computer serial port, or the two devices will not be able communicate.
1. Press
MENU
to view the main menu.
2. Choose
RS-232
, then press
ENTER
.
3. Choose the following options:
– Select
BAUD
and then choose the desired option.
– Choose
BITS
and then select the appropriate option.
– Select
PARITY
and then choose the desired.
– Choose
FLOW-CTRL
and then select the appropriate option.
6. Press
EXIT
until the Main Menu displays.
How do I use Test Script Builder?
Reference
See Section 2 of the Series 2600 Reference Manual for complete details on using
the Test Script Builder.
Step 1: Run Test Script Builder
Run the Test Script Builder program in the Keithley Instruments folder in the usual manner. The
initial startup screen shown in
will be displayed. The main sections of the
screen are briefly described below.
Main Bar
The menu contain following options:
File
– Use this option to control projects and files.
Edit
– Performs a number of editing functions on script files.
Navigate
– Navigates through projects.
Project
– Opens, closes, and builds projects.