2600S-900-01 Rev. C / January 2008
Return to
2-9
Series 2600 System SourceMeter
Section 2: Remote Operation
Note that the reading response will appear in the console window (
).
Alternatively, you can also take and print readings by including the appropriate
measure
command as the argument in the
command. For example, the following commands will
take and print ohms and power readings respectively:
print(smua.measure.r())
print(smua.measure.p())
Again, readings will be displayed in the console window.
Step 7: Turn off output
Send the following command to turn off the output when measurements are complete:
smua.source.output =smua.OUTPUT_OFF
Figure 2-6
Source-measure command sequence in console window
Reset SourceMeter
Select volts function
Set source range to 40V
Set source level to 10V
Set current limit to 10mA
Set measure range to 10mA
Turn on output
Take current reading
Print current reading
Instrument response
Print resistance reading
Instrument response
Print power reading
Instrument response
Turn off output
(2601/2602 version shown)