4-2
Return to
2600S-900-01 Rev. C / January 2008
Section 4: Controlling Multiple Series 2600s (TSP-Link)
Series 2600 System SourceMeter
How do I set up the TSP-Link system?
Up to 16 TSP-Link enabled instruments (e.g., Series 2600 System Series 2600s
TM
instruments)
can be connected together to form a TSP-Link system.
Step 1: Connect the TSP-Link system
Reference
See “Connections” in Section 9 of the Series 2600 Reference Manual for more
information on TSP-Link connections.
A connection example for an expanded system is shown in
. All the TSP-Link-enabled
instruments in the system are daisy-chained together using LAN crossover cables. In this
example, the unit connected to the PC is the Master and the other units are its Slaves. Note that
the Master does not have be assigned as Node 1, nor does it have to be at the end of the node
chain.
Figure 4-1
TSP-Link connections
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
TSPlink
LINE FUSE
SLOWBLOW
3.15A, 250V
CAT I
S
S
LO LO
HI
HI
G
G
G
G
CHANNEL A
MADE IN
U.S.A.
!
!
!
CAT I
S
S
LO
LO
HI
HI
G
G
G
G
CHANNEL B
!
UL
C
US
LISTED
SourceMeter
4ZA4
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
TSPlink
LINE FUSE
SLOWBLOW
3.15A, 250V
CAT I
S
S
LO LO
HI
HI
G
G
G
G
CHANNEL A
MADE IN
U.S.A.
!
!
!
CAT I
S
S
LO
LO
HI
HI
G
G
G
G
CHANNEL B
!
UL
C
US
LISTED
SourceMeter
4ZA4
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
TSPlink
LINE FUSE
SLOWBLOW
3.15A, 250V
CAT I
S
S
LO LO
HI
HI
G
G
G
G
CHANNEL A
MADE IN
U.S.A.
!
!
!
CAT I
S
S
LO
LO
HI
HI
G
G
G
G
CHANNEL B
!
UL
C
US
LISTED
SourceMeter
4ZA4
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
TSPlink
LINE FUSE
SLOWBLOW
3.15A, 250V
CAT I
S
S
LO LO
HI
HI
G
G
G
G
CHANNEL A
MADE IN
U.S.A.
!
!
!
CAT I
S
S
LO
LO
HI
HI
G
G
G
G
CHANNEL B
!
UL
C
US
LISTED
SourceMeter
4ZA4
Node 1
Node 2
Node 3
Node 16
Host
PC
RS-232
GPIB
or
LAN crossover cables
Type: Category 5e or higher.
Length: 3 meters maximum between nodes.
TSP-Link connectors
(2 per instrument)
NOTE
The PC is not needed for
stand-alone systems.
Step 2: Assign node numbers
Reference
See “Initialization” in Section 9 of the Series 2600 Reference Manual for more
information on assigning node numbers.
Each instrument in the TSP-Link system must have a unique node number assigned to it. The
node number for each instrument is stored in its non-volatile memory and will not be lost when the
instrument is turned off.
Complete the following steps to assign a node number.
1. From the front panel, press the
MENU
>
TSPLINK
.
2. Select
NODE
and then use the navigation wheel to assign the node number.
Note: The node number can range from 1 - 64.
For remote programming, the
tsplink.node
attribute is used to set the node number for an
instrument:
tsplink.node = N
where
: N = 1 to 16 The node number of an instrument can be determined by
reading the
tsplink.node
attribute as follows:
nodenumber = tsplink.node