Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
www.keithley.com
Specifications are subject to change without notice.
Page 2 of 7
SPEC-2601_2602 Rev. D / January 2008
Models 2601/2602
System SourceMeter
®
Specifications
Current Programming Accuracy
Range
Programming
resolution
Accuracy (1 year)
23°C + 5°C
+ (% rdg.+amps)
Typical noise
(peak-peak)
0.1Hz-10Hz
100.000nA
1pA
0.06% + 100pA
5pA
1.00000
ȝ
A
10pA
0.03% + 800pA
25pA
10.0000
ȝ
A
100pA
0.03% + 5nA
60pA
100.000
ȝ
A
1nA
0.03% + 60nA
3nA
1.00000mA
10nA
0.03% + 300nA
6nA
10.0000mA
100nA
0.03% + 6
ȝ
A 200nA
100.000mA 1
ȝ
A
0.03% + 30
ȝ
A 600nA
1.00000A 10
ȝ
A
0.05% + 1.8mA
70
ȝ
A
3.00000A 10
ȝ
A
0.06% + 4.0mA
150
ȝ
A
Temperature coefficient (0°C–18°C and 28°C–50°C):
±(0.15 × accuracy specification)/°C.
Maximum output power and source/sink limits:
2
40.4W per channel maximum. ±1.01A at ±40.0V, ±3.03A at
±6.0V, four quadrant source or sink operation.
Current regulation: Line:
0.01% of range.
Load:
±(0.01% of range + 100pA).
Voltage limit/compliance:
4
Bipolar voltage limit (compliance) set with a single value. Minimum value is 10mV.
Accuracy same as voltage source.
Overshoot:
<0.1% typical (step size = 10% to 90% of range, resistive load; see “Current source output settling
time” for additional test conditions).
DC
-3A
-1A
+1A
+3A
+6V
-6V
-40V
+40V
Models 2601 and 2602 I-V capability
4 For sink mode operation (quadrants II and IV), add 10% of compliance range and ±0.02% of limit setting to corresponding
voltage source specification. For 100mV range, add an additional 60mV of uncertainty.