Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
www.keithley.com
Specifications are subject to change without notice.
SPEC-2601_2602 Rev. D / January 2008
Page 3 of 7
Models 2601/2602
System SourceMeter
®
Specifications
Additional Source Specifications
Transient response time:
<70
ȝ
s for the output to recover to 0.1% for a 10% to 90% step change in load.
Voltage source output settling time:
Time required to reach 0.1% of final value, when changing from 10% to 90%
of range, after source level command is processed on a fixed range.
100mV, 1V ranges:
<50
ȝ
s typical.
6V Range:
<100
ȝ
s typical.
40V Range:
<150
ȝ
s typical.
5
Current source output settling time:
Time required to reach 0.1% of final value, when changing from 10% to 90%
of range, after source level command is processed on a fixed range. Values below for Iout · Rload = 1V
unless noted.
3A–10mA ranges:
<80
ȝ
s typical (current less than 2.5A, Rload >1.5
ȍ
).
1mA range:
<100
ȝ
s typical.
100
ȝ
A range:
<150
ȝ
s typical.
10
ȝ
A range:
<500
ȝ
s typical.
1
ȝ
A range:
<2.5ms typical.
100nA range:
<25ms typical.
DC floating voltage:
Output can be floated up to ±250VDC from chassis ground.
Remote sense operating range:
1
Maximum voltage between HI and SENSE HI = 3V.
Maximum voltage between LO and SENSE LO = 3V.
Voltage output headroom:
40V range:
Max. output voltage = 42V – total voltage drop across source leads (maximum 1
ȍ
per source lead).
6V range:
Max. output voltage = 8V – total voltage drop across source leads.
Over temperature protection:
Internally sensed temperature overload puts unit in standby mode.
Voltage source range change overshoot:
Overshoot into a 100k
ȍ
load, 20MHz BW, 300mV typical.
Current source range change overshoot:
<5% of larger range + 300mV/Rload typical.
(see “CURRENT SOURCE OUTPUT SETTLING TIME” for additional test conditions).
5 Add 150
ȝ
s when measuring on the 1A range.