Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
www.keithley.com
Specifications are subject to change without notice.
Page 8 of 8
SPEC-2635_2636 Rev. A / September 2007
Model 2635/2636
System SourceMeter
®
Specifications
Specifications Category
Specifications
Host Interfaces
Computer control interfaces
IEEE-488
IEEE-488.1 compliant. Supports IEEE-488.2 common commands
and status model topology
RS-232
Baud rates from 300bps to 115200bps. Programmable number of
data bits, parity type, and flow control (RTS/CTS hardware or
none). When not programmed as the active host interface, the
SourceMeter can use the RS-232 interface to control other –
instrumentation
10k
Ω
Coil
Resistance
145
Ω ±
10%
Read by firmware
INTERLOCK Pin
(on DIGITAL I/O
connector)
Rear Panel
Chassis
Ground
To output stage
+220V Supply
–220V Supply
Power Supply
100 V to 240 VAC, 50 Hz – 60 Hz (manual setting), 240 VA max
Cooling
Forced air. Side intake and rear exhaust. One side must be
unobstructed when rack mounted
Warranty 1 year
EMC
Conforms to European Union Directive 2004/108/EEC,
EN 61326-1
Safety
Conforms to European Union Directive 73/23/EEC, EN 61010-1,
and UL 61010-1
Dimensions
89 mm high × 213 mm wide × 460 mm deep (31⁄2 in × 83⁄8 in ×
171⁄2 in). Bench Configuration (with handle & feet): 104 mm high ×
238 mm wide × 460 mm deep (41⁄8 in × 93⁄8 in × 171⁄2 in)
Weight
2635: 4.75 Kg (10.4 lbs).
2636: 5.50 Kg (12.0 lbs).
Environment
For indoor use only
Altitude
Maximum 2000 meters above sea level
Operating
0° – 50°C, 70% R.H. up to 35°C. Derate 3% R.H./°C, 35° – 50°C
Storage
– 25°C to 65°C