Series 2600
System SourceMeter
®
Specifications
Reading Buffers:
Non-volatile storage area(s) reserved for measurement data. Reading buffers are arrays of measurement elements.
Each element can hold the following items:
• Measurement
• Source setting (at the time the measurement was taken)
• Measurement status
• Range information
• Timestamp
Two reading buffers are reserved for each SourceMeter channel. Reading buffers can be filled using the front
panel STORE key, and retrieved using the RECALL key or host interface.
Buffer Size, with timestamp and source setting:
>50,000 samples.
Buffer Size, without timestamp and source setting:
>100,000 samples.
Battery Backup:
Lithium-ion battery backup; 30 days of non-volatile storage. Typical battery life is 1 year.
Factory TSP Scripts:
See
www.keithley.com
for Keithley-supported application-specific scripts.
System Expansion:
The TSP-Link expansion interface allows TSP-enabled instruments to trigger and communicate with each other.
See figure below:
Each SourceMeter has two TSP-Link connectors to facilitate chaining instruments together.
• Once SourceMeter instruments are interconnected via TSP-Link, a computer can access all of the resources of
each SourceMeter via the host interface of any SourceMeter.
• A maximum of 16 TSP-Link nodes can be interconnected. Each SourceMeter consumes one TSP-Link node.
TIMER:
Free-running 47-bit counter with 1MHz clock input. Reset each time instrument powers up. Rolls over every 4
years.
Timestamp:
TIMER value automatically saved when each measurement is triggered.
Resolution:
1
ȝ
s.
Accuracy:
100ppm.