Section 3: DUT Test Connections
Series 2600 System SourceMeter® Instruments Reference Manual
Safety shield
A safety shield must be used whenever hazardous voltages (
>
30Vrms, 42Vpeak) will be present in
the test circuit. The safety shield can be metallic or nonmetallic, and must completely surround the
DUT test circuit. A metal safety must be connected to a known safety earth ground and chassis
ground. See the section
later in this section for important safety information on the
use of a metal or nonmetallic enclosure.
Model 2601/2602 safety shield
The maximum output voltage for a Model 2601/2602 channel is 40V, which is considered a non-
hazardous level. However, using two or more Model 2601/2602 voltage sources in a series
configuration can cause test circuit voltage to exceed 42V. For example, the SMUs of two Model
2601/2602 instruments can be connected in series to apply 80V to a DUT (see
).
The connections for the test configuration in
. Use #18AWG
wire or larger for connections to safety earth ground and chassis.
NOTE Floating an SMU may also cause test circuit voltage to exceed 42V
(see
Figure 3-16
HI
LO
HI
LO
DUT
HI
LO
Metal Safety
Shield
SMU A
2601/2602-1
SMU A
2601/2602-2
Safety
Earth
Ground
40V
40V
80V
Safety shield for hazardous voltage using two 2601/2602 channels (>42V)
Figure 3-17
CHANNEL A
LO
G
LO
HI G G G HI
S
S
Model 2601/2602-1
DUT
HI
LO
CHANNEL A
LO
G
LO
HI G G G HI
S
S
Model 2601/2602-2
LO
HI
Safety
Earth
Ground
Chassis
Chassis
Screw
Model 2601/2602-1 connections for test circuit shown in