Series 2600
System SourceMeter
®
Specifications
1. SPEED SPECIFICATIONS
1,2,3
Maximum Sweep Operation Rates (operations per second) for 60Hz (50Hz):
A/D converter
speed
Trigger
origin
Measure to
memory
Measure to
GPIB
Source
measure to
memory
Source
measure to
GPIB
Source
measure
pass/fail to
memory
Source
measure
pass/fail to
GPIB
0.001 NPLC
Internal
10000 (10000)
8000 (8000)
5500 (5500)
3600 (3600)
4900 (4900)
3100 (3100)
0.001 NPLC
Digital I/O
2700 (2650)
2100 (2100)
2300 (2300)
1900 (1875)
2200 (2150)
1800 (1775)
0.01 NPLC
Internal
4000 (3500)
3600 (3200)
2750 (2700)
2300 (2100)
2800 (2500)
2100 (1975)
0.01 NPLC
Digital I/O
1900 (1775)
1600 (1500)
1700 (1600)
1450 (1400)
1600 (1500)
1400 (1325)
0.1 NPLC
Internal
565 (475)
555 (470)
540 (450)
510 (440)
535 (455)
505 (430)
0.1 NPLC
Digital I/O
490 (420)
470 (405)
470 (410)
450 (390)
470 (400)
450 (390)
1.0 NPLC
Internal
59 (49)
59 (49)
58 (49)
58 (48)
58 (49)
58 (48)
1.0 NPLC
Digital I/O
58 (48)
58 (48)
58 (48)
57 (48)
57 (48)
57 (48)
Maximum Single Measurement Rates (operations per second) for 60Hz (50Hz):
A/D converter
speed
Trigger
origin
Measure to
GPIB
Source measure to
GPIB
Source measure pass/fail to GPIB
0.001 NPLC
Internal
1110 (1000)
880 (880)
840 (840)
0.01 NPLC
Internal
950 (900)
780 (760)
730 (710)
0.1 NPLC
Internal
390 (345)
355 (320)
340 (305)
1.0 NPLC
Internal
57 (48)
56 (47)
56 (47)
Maximum measurement range change rate:
>4500/second typical. When changing to or from a range
1A,
maximum rate is >2000/second typical.
Maximum source range change rate:
>400/second, typical.
Maximum source function change rate:
>500/second, typical.
External trigger input:
The digital I/O interface signals can be configured to operate as trigger inputs.
Input latency (time from trigger input to start of measurement or source change):
<150
ȝ
s, typical.
Input jitter:
<100
ȝ
s, typical.
Command processing time:
Maximum time required for the output to begin to change following the receipt of the
smux.source.levelv or smux.source.leveli command. <1ms typical.
1 See the Speed Specifications Test Conditions Appendix in the Series 2600 Reference Manual for more information regarding
test conditions.
2
Exclude current measurement ranges less than 1mA.
3