Section 2: TSP Programming
Series 2600 System SourceMeter® Instruments Reference Manual
Figure 2-3
GPIB Cable
Side View
Side View
Connect one end of the cable to the host
PC and the other end to the SourceMeter.
Both cable connectors are identical.
GPIB cable
NOTE To minimize interference caused by electromagnetic radiation, use
only shielded GPIB cables. Available shielded cables from Keithley
Instruments are the Model 7006 and Model 7007.
The GPIB cable connectors are stackable. For additional non-Series 2600 GPIB instruments in the
test system, daisy-chain a GPIB cable from one instrument to another.
GPIB address
At the factory, the GPIB is set for address value 26. The address value can be set to any address
value between 0 and 30. However, the address cannot conflict with the address assigned to other
instruments in the system.
The GPIB address can be changed from the communications menu. To access the menu, press
the MENU key, select COMMUNICATIONS, and then select GPIB.
The GPIB address is saved in non-volatile memory. The address value will not change when
power is cycled or a reset command (
reset
or *RST) is sent.
RS-232
RS-232 connections
To connect the Series 2600 to the RS-232 interface, use a straight-through RS-232 cable (see
shows the location of the RS-232 connector on the SourceMeter. When
connecting the cable, make sure to tighten the captive screws.
Figure 2-4
Male DB-9 Connector
(connect to SourceMeter)
Pin View
Female DB-9 Connector
(connect to PC)
Pin View
Straight-through RS-232 Cable
RS-232 cable (straight-through)
RS-232 settings
At the factory, the RS-232 is configured as follows: