Section 1: Front and Rear Panel Operation
Series 2600 System SourceMeter
How do I use the buffer?
Reference
See Section 7 of the Series 2600 Reference Manual for more detailed information on
using the buffer.
The instrument has two buffers per channel that can store from 1 to more than 100,000 readings.
Each buffer reading is numbered and can also include the source value and a timestamp.
The following example shows how to store 100 readings, source values, and timestamps in
Channel A, Buffer 1 and recall them from the front panel.
Step 1: Connect the DUT
Connect a 10k
Ω
resistor to the Channel A HI and LO terminals (see
).
Step 2: Set up source and measure functions
Using the procedure described in
“How do I make measurements?” on page 1-12
, set up source
and measure functions:
• Source function: volts
• Source range: 20V (2611/2612/2635/2636) or 40V (2601/2602)
• Source value: 10V
• Measure function: current
• Measure range: auto
Step 3: Configure the buffer
1. From the front panel, press the
STORE
key.
The Count screen displays.
2. Use the
navigation wheel
to select the number of readings and then press
ENTER
.
NOTE You can store up to 100.
3. Choose
CONFIG > STORE
>
CHANA_BUFF
, then the press the navigation wheel to select
Channel A.
4. Select
DEST
>
CHANA_BUFFER1
, and then press
ENTER
.
NOTE The message “
Cant’ change dest buffers” while storing
displays if the
instrument is in store mode.
5. Choose
CHANA
-
BUFF1
, then press the
navigation wheel
to set up Buffer 1.
Complete the following steps to clear the Buffer 1.
1. Select
BUFFER1
>
CLEAR
.
2. Select
YES
.
3. Choose
ELEMENTS
.
4. Select
SRC-VAL
and then choose
ON
to enable the source value.
5. Select
TSTAMP
and then choose
ON
to enable timestamps.
6. Press
EXIT
until the normal display.