SourceMeter
®
Line Specifications
2400, 2400-C, 2410, 2410-C,
2420, 2420-C, 2425, 2425-C,
2430, 2430-C, 2440, 2440-C
ADDITIONAL SOURCE SPECIFICATIONS
TRANSIENT RESPONSE TIME: 30µs minimum for the output to recover
to its spec. following a step change in load.
COMMAND PROCESSING TIME: Maximum time required for the output
to begin to change following the receipt of :SOURce:VOLTage|CURRent
<nrf> command. Autorange On: 10ms. Autorange Off: 7ms.
OUTPUT SETTLING TIME: Time required to reach 0.1% of final value
after command is processed. 100µs typical. Resistive load. 10µA to
100mA range.
OUTPUT SLEW RATE (±30%):
Model 2400:
0.5V/µs, 200V range, 100mA compliance.
0.08V/µs, 20V range, 100mA compliance.
Model 2410:
0.5V/µs, 1000V range, 20mA compliance.
0.08V/µs, 20V range, 100mA compliance.
Model 2420:
0.14V/µs, 60V range, 100mA compliance.
0.08V/µs, 20V range, 100mA compliance.
Model 2425, 2430: 0.25V/µs, 100V range, 100mA compliance.
0.08V/µs, 20V range, 100mA compliance.
Model 2440:
0.25V/µs, 40V range, 100mA compliance.
0.08V/µs, 10V range, 100mA compliance.
DC FLOATING VOLTAGE: Output can be floated up to ±250VDC (Model
2440 ±40VDC) from chassis ground.
REMOTE SENSE: Up to 1V drop per load lead.
COMPLIANCE ACCURACY: Add 0.3% of range and ±0.02% of reading to
base specification.
OVER TEMPERATURE PROTECTION: Internally sensed temperature
overload puts unit in standby mode.
RANGE CHANGE OVERSHOOT: Overshoot into a fully resistive 100k
Ω
load, 10Hz to 1MHz BW, adjacent ranges: 100mV typical, except
20V/200V (20V/60V on Model 2420), 20V/100V on Model 2425 and
2430, range boundary, and Model 2440.
MINIMUM COMPLIANCE VALUE: 0.1% of range.
ADDITIONAL PULSE MODE SOURCE
SPECIFICATIONS
MAXIMUM DUTY CYCLE: 8%, hardware limited, 10A range only. All other
ranges 100%.
MAXIMUM PULSE WIDTH: 5ms from 90% rising to 90% falling edge,
2.5ms 10A range.
MINIMUM PULSE WIDTH: 150µs.
MINIMUM PULSE RESOLUTION: 50µs typical, 70µs max., limited by sys-
tem jitter.
SOURCE ACCURACY: Determined by settling time and source range
specifications.
OUTPUT SETTLING TIME 0.1%:
800µs typ., source I = 10A into 10
Ω
, limited by voltage slew rate.
500µs typ., source I = 10A into 1
Ω
, limited by voltage slew rate.
OUTPUT SLEW RATE:
Voltage (10
Ω
load):
0.25V/µs ±30% on 100V range
0.08V/µs ±30% on 20V range, 10A range.
Current (0
Ω
load):
0.25A/µs ±30% on 100V range
0.08A/µs ±30% on 20V range, 10A range.
VOLTAGE PROGRAMMING ACCURACY (Local or Remote Sense)
ACCURACY (1 Year)
NOISE
PROGRAMMING
23°C ±5°C
(peak-peak)
MODEL
RANGE
RESOLUTION
±(% rdg. + volts)
0.1Hz – 10Hz
ALL:
200.000 mV
5 µV
0.02% + 600 µV
5 µV (10µV 2430, 2440)
2.00000
V
50 µV
0.02% + 600 µV
50 µV
2400 Only:
20.0000
V
500 µV
0.02% + 2.4 mV
500 µV
200.000
V
5 mV
0.02% + 24 mV
5 mV
2410 Only:
20.0000
V
500 µV
0.02% + 2.4 mV
500 µV
1000.00
V
50 mV
0.02% + 100 mV
20 mV
2420 Only:
20.0000
V
500 µV
0.02% + 2.4 mV
500 µV
60.0000
V
1.5 mV
0.02% + 7.2 mV
1.5 mV
2425 and
20.0000
V
500 µV
0.02% + 2.4 mV
500 µV
2430 Only: 100.0000
V
2.5 mV
0.02% + 12 mV
2.5 mV
2440 Only:
10.0000
V
500 µV
0.02% + 1.2 mV
250 µV
40.0000
V
5 mV
0.02% + 4.8 mV
1 mV
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.15
×
accuracy specification)/°C.
MAX. OUTPUT POWER: 22W (66W for Model 2420, 110W for Model 2425 and 2430, 55W for
Model 2440), four quadrant source or sink operation.
SOURCE/SINK LIMITS: Model 2400: ±21V @ ±1.05A, ±210V @ ±105 mA.
Model 2410: ±21V @ ±1.05A, ±1100V @ ±21 mA.
Model 2420: ±21V @ ±3.15A, ±63V @ ±1.05A.
Model 2425: ±21V @ ±3.15A, ±105V @ ±1.05A.
Model 2430: ±105V @ ±1.05A, ±105V @ ±10.5A (pulse mode only).
Model 2440: ±10.5V @ ±5.25A, ±42V @ ±1.05A.
VOLTAGE REGULATION: Line: 0.01% of range. Load: 0.01% of range + 100µV.
NOISE 10Hz–1MHz (p-p): 10mV (50mV typ., Models 2430 and 2440).
OVER VOLTAGE PROTECTION: User selectable values, 5% tolerance. Factory default = none.
CURRENT LIMIT: Bipolar current limit (compliance) set with single value. Min. 0.1% of range.
OVERSHOOT: <0.1% typical (full scale step, resistive load, 10mA range).
CURRENT PROGRAMMING ACCURACY (Local or Remote Sense)
ACCURACY (1 Year)
3
NOISE
PROGRAMMING
23°C ±5°C
(peak-peak)
MODEL
RANGE
RESOLUTION
±(% rdg. + amps)
0.1Hz – 10Hz
2400, 2410:
1.00000 µA
50 pA
0.035% + 600 pA
5 pA
ALL:
10.0000 µA
500 pA
0.033% +
2 nA
5 nA
100.000 µA
5 nA
0.031% + 20 nA
50 nA
1.00000 mA
50 nA
0.034% + 200 nA
500 nA
2400, 2420,2425, 10.0000 mA
500 nA
0.045% +
2 µA
50 µA
2430, 2440:
2410 Only:
20.0000 mA
500 nA
0.045% +
4 µA
200 nA
ALL:
100.000 mA
5 µA
0.066% + 20 µA
1 µA
2400, 2410:
1.00000
A
2
50 µA
0.27% + 900 µA
100 µA
2420, 2425,
1.00000
A
50 µA
0.067% + 900 µA
50 µA
2430, 2440:
2420, 2425:
3.00000
A
2
150 µA
0.059% + 2.7 mA
150 µA
2430:
3.00000
A
2
/
500 µA
0.059% + 2.8 mA
300 µA
10.00000
A
4
2440:
5.00000
A
50 µA
0.10% + 5.4 mA
500 µA
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.15
×
accuracy specification)/°C.
MAX. OUTPUT POWER: 22W (66W for Model 2420, 110W for Model 2425 and 2430, 55W for
Model 2440), four quadrant source or sink operation.
SOURCE/SINK LIMITS: Model 2400: ±1.05A @ ±21V, ±105 mA @ ±210V.
Model 2410: ±1.05A @ ±21V, ±21 mA @ ±1100V.
Model 2420: ±3.15A @ ±21V, ±1.05A @ ±63V.
Model 2425: ±3.15A @ ±21V, ±1.05A @ ±105V.
Model 2430: ±1.05A @ ±105V, ±10.5A @ ±105V (pulse mode only).
Model 2440: ±5.25A @ ±10.5V, ±1.05A @ ±42V.
CURRENT REGULATION: Line: 0.01% of range. Load: 0.01% of range (except Model 2440 5A
range 0.05%) + 100pA.
VOLTAGE LIMIT: Bipolar voltage limit (compliance) set with single value. Min. 0.1% of range.
OVERSHOOT: <0.1% typical (1mA step, RL = 10k
Ω
, 20V range for Model 2400, 2410, 2420, 2425,
2430), (10V range for Model 2440).
SOURCE SPECIFICATIONS
1
1 2400, 2410 Only: Specifications valid for continuous output currents below
105mA. For operation above 105mA continuous for > 1 minute, derate accuracy
10%/35mA above 105mA.
2 Full operation (1A) regardless of load to 30°C (50°C for Model 2420, 2440). Above
30°C (50°C for Model 2420, 2440) ambient, derate 35mA/°C and prorate 35mA/
Ω
load. 4-wire mode. For current sink operation on 1A, 3A, or 5A ranges, maximum
continuous power is limited to approximately 1/2 rated power or less, depending
on current, up to 30°C ambient. See power equations in the User’s Manual to cal-
culate allowable duty cycle for specific conditions.
3 For sink mode, 1µA to 100mA range, accuracy is:
Model 2400: ±(0.15% + offset*4)
Models 2410, 2420, 2425, 2430, 2440: ±(0.5% + offset*3)
For 1A range, accuracy is:
Model 2400: ±(1.5% + offset*8)
Models 2410, 2420, 2425, 2430, 2440: ±(1.5% + offset*3)
4 10A range only in pulse mode. Limited to 2.5ms pulse width maximum. 10% duty
cycle maximum.
Rev. A
5/17/01
Summary of Contents for SourceMeter 3A 2420
Page 2: ...Model 24203A SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 15: ...1 Performance Verification...
Page 31: ...2 Calibration...
Page 53: ...3 Routine Maintenance...
Page 56: ...3 4 Routine Maintenance...
Page 57: ...4 Troubleshooting...
Page 70: ...4 14 Troubleshooting...
Page 71: ...5 Disassembly...
Page 78: ...5 8 Disassembly...
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ...6 Replaceable Parts...
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ...A Specifications...
Page 112: ...A 6 Specifications...
Page 113: ...B Command Reference...
Page 124: ...B 12 Command Reference...
Page 125: ...C Calibration Programs...