SourceMeter
®
Line Specifications
RESISTANCE MEASUREMENT ACCURACY (Local or Remote Sense)
DEFAULT
DEFAULT
NORMAL
ENHANCED
TEST
TEST CURRENT
ACCURACY (23°C ±5°C)
ACCURACY (23°C ±5°C)
5
DEFAULT
CURRENT
2420, 2425,
1 YEAR, ±(% rdg. + ohms)
1 YEAR, ±(% rdg. + ohms)
RANGE
RESOLUTION
2400, 2410
2430, 2440
2400
2410
2420, 2425, 2430, 2440
2400
<0.20000
Ω
4
–
–
–
Source I
ACC
+ Meas.V
ACC
Source I
ACC
+ Meas.V
ACC
Source I
ACC
+ Meas.V
ACC
Source I
ACC
+ Meas.V
ACC
2.00000
Ω
4
10 µ
Ω
–
1
A
Source I
ACC
+ MeasV
ACC
Source I
ACC
+ Meas.V
ACC
0.17% +0.0003
Ω
Source I
ACC
+ Meas.V
ACC
20.0000
Ω
100 µ
Ω
100 mA
100 mA
0.10% + 0.003
Ω
0.11% + 0.006
Ω
0.10% + 0.003
Ω
0.07% + 0.001
Ω
200.000
Ω
1 m
Ω
10 mA
10 mA
0.08% + 0.03
Ω
0.09% +
0.1
Ω
0.08% +
0.03
Ω
0.05% + 0.01
Ω
2.00000 k
Ω
10 m
Ω
1 mA
1 mA
0.07% +
0.3
Ω
0.08% +
0.6
Ω
0.07% +
0.3
Ω
0.05% +
0.1
Ω
20.0000 k
Ω
100 m
Ω
100 µA
100 µA
0.06% +
3
Ω
0.07% +
6
Ω
0.06% +
3
Ω
0.04% +
1
Ω
200.000 k
Ω
1
Ω
10 µA
10 µA
0.07% +
30
Ω
0.07% +
60
Ω
0.07% +
30
Ω
0.05% +
10
Ω
2.00000 M
Ω
8
10
Ω
1 µA
1 µA
0.11% +
300
Ω
0.12% +
600
Ω
0.11% +
300
Ω
0.05% +
100
Ω
20.0000 M
Ω
9
100
Ω
1 µA
1 µA
0.11% +
1 k
Ω
0.12% +
2.4 k
Ω
0.11% +
1 k
Ω
0.05% +
500
Ω
200.000 M
Ω
4
1 k
Ω
100 nA
–
0.66% +
10 k
Ω
0.66% +
24 k
Ω
Source I
ACC
+ Meas.V
ACC
0.35% +
5 k
Ω
>200.000 M
Ω
4
–
–
–
Source I
ACC
+ Meas.V
ACC
Source I
ACC
+ Meas.V
ACC
Source I
ACC
+ Meas.V
ACC
Source I
ACC
+ Meas.V
ACC
VOLTAGE MEASUREMENT ACCURACY (Local or Remote Sense)
ACCURACY (1 Year)
DEFAULT
INPUT
23°C ±5°C
MODEL
RANGE
RESOLUTION
RESISTANCE
±(% rdg. + volts)
ALL:
200.000 mV
1 µV
>10 G
Ω
0.012% + 300 µV
2.00000
V
10 µV
>10 G
Ω
0.012% + 300 µV
2400:
20.0000
V
100 µV
>10 G
Ω
0.015% + 1.5 mV
200.000
V
1 mV
>10 G
Ω
0.015% + 10 mV
2410:
20.0000
V
100 µV
>10 G
Ω
0.015% +
1 mV
1000.00
V
10 mV
>10 G
Ω
0.015% + 50 mV
2420, 2425, 2430:
20.0000
V
100 µV
>10 G
Ω
0.015% +
1 mV
2420:
60.0000
V
1 mV
>10 G
Ω
0.015% +
3 mV
2430:
100.000
V
1 mV
>10 G
Ω
0.015% +
5 mV
2440:
10.0000
V
100 µV
>10 G
Ω
0.015% + 750 µV
40.0000
V
1 mV
>10 G
Ω
0.015% +
3 mV
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.15
×
accuracy specification)/°C.
CURRENT MEASUREMENT ACCURACY (Local or Remote Sense)
ACCURACY (1 Year)
DEFAULT
VOLTAGE
23°C ±5°C
MODEL
RANGE
RESOLUTION
BURDEN
3
±(% rdg. + amps)
2400, 2410:
1.00000 µA
10 pA
< 1mV
0.029% + 300 pA
All:
10.0000 µA
100 pA
< 1mV
0.027% + 700 pA
100.000 µA
1 nA
< 1mV
0.025% +
6 nA
1.00000 mA
10 nA
< 1mV
0.027% + 60 nA
2400, 2420, 2425, 2430, 2440:
10.0000 mA
100 nA
< 1mV
0.035% + 600 nA
2410:
20.0000 mA
100 nA
< 1mV
0.035% + 1.2 µA
All:
100.000 mA
1 µA
< 1mV
0.055% +
6 µA
2400, 2410:
1.00000
A
10 µA
< 1mV
0.22% + 570 µA
2420, 2425, 2430, 2440:
1.00000
A
10 µA
< 1mV
0.060% + 570 µA
2420, 2425:
3.00000
A
10 µA
< 1mV
0.052% + 1.71 mA
2430:
3.00000
A
10 µA
< 1mV
0.052% + 1.71 mA
10.0000
A
6
2440:
5.00000
A
10 µA
< 1mV
0.10% + 3.42 mA
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.10
×
accuracy specification)/°C.
MEASURE SPECIFICATIONS
1, 2, 7
1
Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200mV, 1A, 10A
ranges, add 0.05%. For 0.01 PLC, add 0.05% of range to offset specifications, except 200mV, 1A, 10A
ranges, add 0.5%.
2
Accuracies apply to 2- or 4-wire mode when properly zeroed.
3
4-wire mode.
4
Manual ohms only – except 2420, 2425, 2430, 2440 for 2
Ω
range and 2410 or 2400 for 200M
Ω
range.
5
Source readback enabled, offset compensation ON. Also available on 2410, 2420, 2425, 2430, and 2440
with similar accuracy enhancement.
6
10A range only in pulse mode.
7
In pulse mode, limited to 0.1 PLC measurement.
8
Except 2440; default test current is 5µA.
9
Except 2440; default test current is 0.5µA.
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.15
×
accuracy
specification)/°C.
SOURCE I MODE, MANUAL OHMS: Total uncertainty = I source accuracy
+ V measure accuracy (4-wire remote sense).
SOURCE V MODE, MANUAL OHMS: Total uncertainty = V source accuracy
+ I measure accuracy (4-wire remote sense).
6-WIRE OHMS MODE: Available using active ohms guard and guard
sense. Max. Guard Output Current: 50mA (except 1A range). Accuracy
is load dependent. Refer to White Paper no. 2033 for calculation
formula.
GUARD OUTPUT IMPEDANCE: <0.1
Ω
in ohms mode.
Rev. A
5/17/01
CONTACT CHECK
SPECIFICATIONS
SPEED: 350µs for verification and notification.
CONTACT CHECK:
2
Ω
15
Ω
50
Ω
No contact
check failure
<1.00
Ω
<13.5
Ω
<47.5
Ω
Always contact
check failure
>3.00
Ω
>16.5
Ω
>52.5
Ω
2400, 2400-C, 2410, 2410-C,
2420, 2420-C, 2425, 2425-C,
2430, 2430-C, 2440, 2440-C
Summary of Contents for SourceMeter 3A 2420
Page 2: ...Model 24203A SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 15: ...1 Performance Verification...
Page 31: ...2 Calibration...
Page 53: ...3 Routine Maintenance...
Page 56: ...3 4 Routine Maintenance...
Page 57: ...4 Troubleshooting...
Page 70: ...4 14 Troubleshooting...
Page 71: ...5 Disassembly...
Page 78: ...5 8 Disassembly...
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ...6 Replaceable Parts...
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ...A Specifications...
Page 112: ...A 6 Specifications...
Page 113: ...B Command Reference...
Page 124: ...B 12 Command Reference...
Page 125: ...C Calibration Programs...