B-10
Command Reference
Detecting calibration step completion
When sending remote calibration commands, you must wait until the instrument completes
the current operation before sending another command. You can use either *OPC? or *OPC to
help determine when each calibration step is completed.
Using the *OPC? query
With the *OPC? (operation complete) query, the instrument will place an ASCII 1 in the out-
put queue when it has completed each step. To determine when the OPC response is ready, do
the following:
1.
Repeatedly test the MAV (Message Available) bit (bit 4) in the status byte and wait until
it is set. (You can request the status byte by using the *STB? query.)
2.
When MAV is set, a message is available in the output queue, and you can read the output
queue and test for an ASCII 1.
3.
After reading the output queue, repeatedly test MAV again until it clears. At this point,
the calibration step is completed.
Using the *OPC command
The *OPC (operation complete) command can also be used to detect the completion of each
calibration step. In order to use *OPC to detect the end of each calibration step, do the following:
1.
Enable operation complete by sending *ESE 1. This command sets the OPC (operation
complete bit) in the standard event enable register, allowing operation complete status
from the standard event status register to set the ESB (event summary bit) in the status
byte when operation complete is detected.
2.
Send the *OPC command immediately following each calibration command. For example:
:CAL:PROT:SENS 2;*OPC
Note that you must include the semicolon (;) to separate the two commands, and that the
*OPC command must appear on the same line as the command.
3.
After sending a calibration command, repeatedly test the ESB (Event Summary) bit (bit
5) in the status byte until it is set. (Use *STB? to request the status byte.)
4.
Once operation complete has been detected, clear OPC status using one of two methods:
(1) use the *ESR? query, then read the response to clear the standard event status register,
or (2) send the *CLS command to clear the status registers. Note that sending *CLS will
also clear the error queue and operation complete status.
Summary of Contents for SourceMeter 3A 2420
Page 2: ...Model 24203A SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 15: ...1 Performance Verification...
Page 31: ...2 Calibration...
Page 53: ...3 Routine Maintenance...
Page 56: ...3 4 Routine Maintenance...
Page 57: ...4 Troubleshooting...
Page 70: ...4 14 Troubleshooting...
Page 71: ...5 Disassembly...
Page 78: ...5 8 Disassembly...
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ...6 Replaceable Parts...
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ...A Specifications...
Page 112: ...A 6 Specifications...
Page 113: ...B Command Reference...
Page 124: ...B 12 Command Reference...
Page 125: ...C Calibration Programs...