Calibration 2-5
4.
Once the correct password is displayed, press the ENTER key. If the password was cor-
rectly entered, the following message will be displayed:
CALIBRATION UNLOCKED
Calibration can now be executed.
5.
Press EXIT to return to normal display. Calibration will be unlocked and assume the
states summarized in Table 2-2. Attempts to change any of the settings listed below with
calibration unlocked will result in an error +510, "Not permitted with cal un-locked."
NOTE
With calibration unlocked, the sense function and range track the source function and
range. That is, when :SOUR:FUNC is set to VOLT, the :SENS:FUNC setting will be
'VOLT:DC'. When :SOUR:FUNC is set to CURR, the :SENS:FUNC setting will be
'CURR:DC'. A similar command coupling exists for :SOUR:VOLT:RANG/
:SENS:VOLT:RANG and SOUR:CURR:RANG:SENS:CURR:RANG.
Unlocking calibration by remote
To unlock calibration via remote, send the following command:
:CAL:PROT:CODE '<password>'
For example, the following command uses the default password:
:CAL:PROT:CODE 'KI002420'
Table 2-2
Calibration unlocked states
Mode
State
Equivalent remote command
Concurrent Functions
Sense Function
Sense Volts NPLC
Sense Volts Range
Sense Current NPLC
Sense Current Range
Filter Count
Filter Control
Filter Averaging
Source V Mode
Volts Autorange
Source I Mode
Current Autorange
Autozero
Trigger Arm Count
Trigger Arm Source
Trigger Count
Trigger Source
OFF
Source
1.0
Source V
1.0
Source I
10
REPEAT
ON
FIXED
OFF
FIXED
OFF
ON
1
Immediate
1
Immediate
:SENS:FUNC:CONC OFF
:SENS:FUNC <source_ function>
:SENS:VOLT:NPLC 1.0
:SENS:VOLT:RANG <source_V_range>
:SENS:CURR:NPLC 1.0
:SENS:CURR:RANG <source_ I_ range>
:SENS:AVER:COUN 10
:SENS:AVER:TCON REPeat
:SENS:AVER:STAT ON
:SOUR:VOLT:MODE FIXED
:SOUR:VOLT:RANG:AUTO OFF
:SOUR:CURR:MODE FIXED
:SOUR:CURR:RANG:AUTO OFF
:SYST:AZERO ON
:ARM:COUNT 1
:ARM:SOUR IMMediate
:TRIG:COUNT 1
:TRIG:SOUR IMMediate
Summary of Contents for SourceMeter 3A 2420
Page 2: ...Model 24203A SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 15: ...1 Performance Verification...
Page 31: ...2 Calibration...
Page 53: ...3 Routine Maintenance...
Page 56: ...3 4 Routine Maintenance...
Page 57: ...4 Troubleshooting...
Page 70: ...4 14 Troubleshooting...
Page 71: ...5 Disassembly...
Page 78: ...5 8 Disassembly...
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ...6 Replaceable Parts...
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ...A Specifications...
Page 112: ...A 6 Specifications...
Page 113: ...B Command Reference...
Page 124: ...B 12 Command Reference...
Page 125: ...C Calibration Programs...