2-10
Calibration
Table 2-3
Front panel voltage calibration
*Use EDIT and RANGE keys to select source range.
**Multimeter reading used in corresponding calibration step. See procedure.
Step 3: Current calibration
Perform the steps below for each current range, using Table 2-4 as a guide.
1.
Connect the Model 2420 to the digital multimeter, as shown in Figure 2-2. Select the
multimeter DC current measurement function.
Source
range*
Source voltage
Multimeter voltage
reading**
0.2V
+200.00mV
+000.00mV
-200.00mV
-000.00mV
___________ mV
___________ mV
___________ mV
___________ mV
2V
+2.0000V
+0.0000V
-2.0000V
-0.0000V
___________ V
___________ V
___________ V
___________ V
20V
+20.000V
+00.000V
-20.000V
-00.000V
___________ V
___________ V
___________ V
___________ V
60V
+60.000V
+00.000V
-60.000V
-00.000V
___________ V
___________ V
___________ V
___________ V
Model 2420
Digital Multimeter
Input LO
Amps
2420 3A SourceMeter
250V
PEAK
5V
PEAK
HI
LO
OUTPUT
75V
PEAK
75V
PEAK
EDIT
TOGGLE
POWER
RANGE
INPUT/
OUTPUT
4- WIRE
SENSE
DISPLAY
ON/OFF
TERMINALS
FRONT/
REAR
AUTO
RANGE
EXIT
ENTER
CONFIG MENU
SWEEP
TRIG
REL
LOCAL
FILTER
LIMIT
DIGITS SPEED
V
Ω
MEAS
I
FCTN
V
I
SOURCE
2
3
0
1
6
7
8
9
4
+/-
5
STORE RECALL
EDIT
Figure 2-2
10
µ
A to 1A range current
calibration test connections
Summary of Contents for SourceMeter 3A 2420
Page 2: ...Model 24203A SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 15: ...1 Performance Verification...
Page 31: ...2 Calibration...
Page 53: ...3 Routine Maintenance...
Page 56: ...3 4 Routine Maintenance...
Page 57: ...4 Troubleshooting...
Page 70: ...4 14 Troubleshooting...
Page 71: ...5 Disassembly...
Page 78: ...5 8 Disassembly...
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ...6 Replaceable Parts...
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ...A Specifications...
Page 112: ...A 6 Specifications...
Page 113: ...B Command Reference...
Page 124: ...B 12 Command Reference...
Page 125: ...C Calibration Programs...