2-22
Calibration
NOTE
Calibration will be temporary unless you send the SAVE command. Also, calibra-
tion data will not be saved if (1) calibration is locked, (2) invalid data exists, or
(3) all steps were not completed.
Step 6: Lock out calibration
To lock out further calibration, send the following command after completing the calibration
procedure:
:CAL:PROT:LOCK
Single-range calibration
Normally, the complete calibration procedure should be performed to ensure that the entire
instrument is properly calibrated. In some instances, however, you may want to calibrate only
certain ranges. To do so, complete the entire procedure only for the range(s) to be calibrated.
Keep in mind, however, that you must complete all parameter steps for each source or sense
range. Also be sure to set calibration dates and save calibration after calibrating the desired
range(s).
Summary of Contents for SourceMeter 3A 2420
Page 2: ...Model 24203A SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 15: ...1 Performance Verification...
Page 31: ...2 Calibration...
Page 53: ...3 Routine Maintenance...
Page 56: ...3 4 Routine Maintenance...
Page 57: ...4 Troubleshooting...
Page 70: ...4 14 Troubleshooting...
Page 71: ...5 Disassembly...
Page 78: ...5 8 Disassembly...
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ...6 Replaceable Parts...
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ...A Specifications...
Page 112: ...A 6 Specifications...
Page 113: ...B Command Reference...
Page 124: ...B 12 Command Reference...
Page 125: ...C Calibration Programs...