background image

 

PA-742 Rev. B / 10-02

 

Model STP-36CJC  Screw Terminal Connector

 

Packing List

 

Keithley Instruments, Inc.

 

28775 Aurora Road
Cleveland, Ohio 44139
(440) 248-0400
Fax: (440) 248-6168
www.keithley.com

 

Introduction

 

This document contains information necessary to successfully install your hardware.

 

NOTE

 

Only used with analog connector on KPCI-3108/7 boards.

 

WARNING

 

The STP-36CJC is not intended for use in circuits carrying voltages in excess of 
30V RMS, 42.4V peak, or 60VDC.

 

General definitions

 

The types of product users are:

 

Responsible body

 

 is the individual or group responsible for the use and maintenance of equipment, and for ensuring that oper-

ators are adequately trained.

 

Operators

 

 use the product for its intended function. They must be trained in electrical safety procedures and proper use of the

instrument. They must be protected from electric shock and contact with hazardous live circuits.

 

Maintenance personnel

 

 perform routine procedures on the product to keep it operating, for example, setting the line voltage

or replacing consumable materials. Maintenance procedures are described in the manual. The procedures explicitly state if the
operator may perform them. Otherwise, they should be performed only by service personnel.

 

Service personnel

 

 are trained to work on live circuits, and perform safe installations and repairs of products. Only properly

trained service personnel may perform installation and service procedures.

The 

 

 

 

symbol on an instrument indicates that the user should refer to the operating instructions located in the manual.

The 

 symbol on an instrument shows that it can source or measure 1000 volts or more, including the combined effect of

normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.

The 

 

WARNING

 

 heading in a manual explains dangers that might result in personal injury or death. Always read the associated

information very carefully before performing the indicated procedure.

The 

 

CAUTION

 

 heading in a manual explains hazards that could damage the instrument. Such damage may invalidate the warranty.

 

Descriptions

 

The screw terminals on the STP-36CJC screw terminal connector let you connect field wiring to the analog connector on the
KPCI-3108/7 board using a CAB-1284CC cable. The screw terminals are labeled from 1 to 36 and correspond directly to the
functions of the pins on the main analog I/O connector on the KPCI-3108/7 board. For example, if pin 24 is assigned to the
analog function OP0, use screw terminal 24 to attach hardware to output 0. Refer to the KPCI-3108/7 Series User’s Manual  for
detailed wiring information on connecting the STP-36CJC.

!

Summary of Contents for STP-36CJC

Page 1: ...nnel Service personnel are trained to work on live circuits and perform safe installations and repairs of products Only properly trained service personnel may perform installation and service procedures The symbol on an instrument indicates that the user should refer to the operating instructions located in the manual The symbol on an instrument shows that it can source or measure 1000 volts or mo...

Page 2: ...ted instrumentation Although some instru ments and accessories would normally be used with non hazardous voltages there are situations where hazardous conditions may be present This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury Read the operating information carefully before using the pr...

Page 3: ...es the use of a lid interlock Instrumentation and accessories shall not be connected to humans Maintenance and service Inspect the connecting cables test leads and jumpers for possible wear cracks or breaks before each use Before performing any maintenance disconnect the line cord and all test cables Cleaning Keep the connections free of contaminants such as dirt oil etc in order to maintain maxim...

Page 4: ...inal Pin Terminal 1 1 13 13 25 25 2 2 14 14 26 26 3 3 15 15 27 27 4 4 16 16 28 28 5 5 17 17 29 29 6 6 18 18 30 30 7 7 19 19 31 31 8 8 20 20 32 32 9 9 21 21 33 33 10 10 22 22 34 34 11 11 23 23 35 35 12 12 24 24 36 36 Figure 1 Pin connections J2 J3 WARNING NOT INTENDED FOR USE WITH CONDUCTORS CARRYING VOLTAGES IN EXCESS OF 30VRMS 60VDC R1 R2 36 STP 36CJC 18 36 1 19 CT3375 35 34 33 32 31 30 29 28 27 ...

Page 5: ...ore if the STP 36CJC is configured for differen tial mode and thermocouples are connected you must connect a 10 to 100kΩ bias resistor between the thermocouple and AGND pin 17 or 18 2 Single ended mode Set jumper J3 in place Remove jumper J2 Figure 3 The STP 36CJC is now configured to provide a CJC voltage to pin 33 Channel 0 on the 3108 analog connector WARNING Do NOT connect any external devices...

Page 6: ... using the STP 36CJC for any kind of temperature measurement it is highly recommended that you enclose the STP 36CJC in an insulated case not included Doing so will greatly reduce any temperature gradients across the board ensuring a more accurate reading Figure 5 Board dimensions J3 J2 2 840 72 1mm 1 250 31 8mm 0 780 19 8mm 0 200 5 1mm 4 820 122 4mm 5 220 132 6mm 0 147 Ø 4 Places 3 7mm STP 36 CJC...