background image

 

Models 2601B, 2602B and 2604B System SourceMeter Instrument Specifications 

 

Specifications are subject to change without notice 

SPEC-2602B Rev. B / February 2014 

5  of 13 

Voltage output headroom 

40 V range 

 

Maximum output voltage = 42 V – (total voltage drop across source leads).  
Maximum 1 

Ω source lead. 

6 V range 

 

Maximum output voltage = 8 V – (total voltage drop across source leads).  
Maximum 1 

Ω source lead. 

Over-temperature protection 

Internally sensed temperature overload puts the instrument in standby mode 

Limit/compliance 

Bipolar limit (compliance) set with a single value 

Voltage:

10

 

 

Minimum value is 10 mV; accuracy is the same as voltage source 

Current:

11

 

 

Minimum value is 10 nA; accuracy is the same as current source 

Voltage source output 
settling time 

Time required to reach within 0.1 % of final value after source level command is 
processed on a fixed range. 

Voltage range 

Settling time  

100 mV 

< 50 µs  

1 V 

< 50 µs 

10 V 

< 110 µs 

40 V

12

 

< 150 µs 

Current source output 
settling time 

Time required to reach within 0.1 % of final value after source level command is 
processed on a fixed range 

 

Values below for I

out

 × R

load

 = 1 V 

Current range 

Settling time 

3 A  

< 80 µs (Current < 2.5 A, R

load

 > 2 

Ω) 

1 A to 10 mA 

< 80 µs (R

load

 > 6 

Ω) 

1 mA 

< 100 µs 

100 µA 

< 150 µs 

10 µA 

< 500 µs 

1 µA 

< 2 ms 

100 nA 

< 20 ms 

                                              

10

 For sink operation (quadrants II and IV) without sink mode enabled, add 10 % of compliance range and ±0.02 % of limit settling to the 

corresponding voltage source accuracy specifications. For 100 mV range add an additional 60 mV of uncertainty. Specifications apply with 
sink mode enabled. 

11

 For sink operation (quadrants II and IV) without sink mode enabled, add 0.06 % of limit range to the corresponding current limit accuracy 

specifications. Specifications apply with sink mode enabled. 

12

 Add 150 µs when measuring on the 1 A range. 

Summary of Contents for System SourceMeter 2601B

Page 1: ...and measurement accuracies are specified at the Models 2601B 2602B and 2604B terminals under these conditions 1 23 C 5 C 70 percent relative humidity 2 After a two hour warm up period 3 Speed normal...

Page 2: ...pA 0 03 5 nA 60 pA 10 pA 0 025 1 5 nA 100 A 2 nA 0 03 60 nA 3 nA 100 pA 0 02 25 nA 1 mA 20 nA 0 03 300 nA 6 nA 1 nA 0 02 200 nA 10 mA 200 nA 0 03 6 A 200 nA 10 nA 0 02 2 5 A 100 mA 2 A 0 03 30 A 600...

Page 3: ...ibe the typical performance of the Models 2601B 2602B and 2604B PULSE CHARACTERISTICS Pulse region specifications Pulse region specifications Region quadrant diagram Region maximums Maximum pulse widt...

Page 4: ...HARACTERISTICS Noise 10 Hz to 20 MHz 20 mV peak peak 3 mV RMS 6 V range Transient response time 70 s for the output to recover to within 0 1 for a 10 to 90 step change in load Overshoot Voltage 0 1 10...

Page 5: ...nal value after source level command is processed on a fixed range Voltage range Settling time 100 mV 50 s 1 V 50 s 10 V 110 s 40 V 12 150 s Current source output settling time Time required to reach...

Page 6: ...quired to reach within 0 1 of final value after source level command is processed on a fixed range Values below for Vout 1 V Current range Settling time 1 mA 100 s Input impedance 10 G ADDITIONAL CHAR...

Page 7: ...ge Settling time 3 A and 1A 120 s Rload 2 100 mA and 10 mA 100 s 1 mA 3 ms 100 A 3 ms 10 A 230 ms 1 A 230 ms Capacitor leakage performance Using HIGH C scripts 19 200 ms 50 nA Load 5 F in parallel wit...

Page 8: ...01 NPLC Digital I O 3500 3100 3400 3000 3000 2700 2900 2600 4150 3650 3800 3400 0 1 NPLC Internal 580 480 560 470 550 465 550 460 560 470 545 460 0 1 NPLC Digital I O 550 460 550 460 540 450 540 450 5...

Page 9: ...maximum rate is 190 second Maximum source function change rate 1000 per second Command processing time 1 ms Maximum time required for the output to begin to change after receiving the smua source leve...

Page 10: ...ts for example branching looping and math Able to execute high speed test scripts stored in memory without host intervention Minimum user memory available 16 MB approximately 250 000 lines of TSP code...

Page 11: ...ink expansion interface allows TSP enabled instruments to trigger and communicate with each other See the figure below Each Model 2601B and 2602B has two TSP Link connectors to make it easier to conne...

Page 12: ...or higher LAN crossover cable 9 84 ft 3 m maximum between each TSP enabled instrument USB Control Rear USB 2 0 Device USB TMC488 protocol USB File System Front USB 2 0 Host Mass storage class device...

Page 13: ...current flowing out of digital I O bit 960 A Maximum sink current at maximum logic low voltage 0 7 5 0 mA Absolute maximum sink current flowing into digital I O pin 11 mA 5 V power supply pins Limited...

Reviews: