background image

Models 2601B, 2602B and 2604B System SourceMeter

®

 Instrument Specifications 

 

Specifications are subject to change without notice 

8 of 13 

SPEC-2602B Rev. B / February 2014 

MEASUREMENT SPEED CHARACTERISTICS

20,21

 

Maximum sweep operation rates (operations per second) for 60 Hz (50 Hz): 

A/D 
converter 
speed  

Trigger 
origin 

Measure to 
memory 
 

(using user 
scripts) 

Measure to 
GPIB 
 

(using user 
scripts) 

Source 
measure to 
memory 

(using user 
scripts) 

Source 
measure to 
GPIB 

(using user 
scripts) 

Source 
measure to 
memory 

(using sweep 
API) 

Source 
measure to 
GPIB 

(using sweep 
API) 

0.001 NPLC 

Internal 

20000 
(20000) 

9800 
(9800) 

7000 
(7000) 

6200 
(6200) 

12000 
(12000) 

5900 
(5900) 

0.001 NPLC 

Digital I/O 

8100 
(8100) 

7100 
(7100) 

5500 
(5500) 

5100 
(5100) 

11200 
(11200) 

5700 
(5700) 

0.01 NPLC 

Internal 

4900 
(4000) 

3900 
(3400) 

3400 
(3000) 

3200 
(2900) 

4200 
(3700) 

4000 
(3500) 

0.01 NPLC 

Digital I/O 

3500 
(3100) 

3400 
(3000) 

3000 
(2700) 

2900 
(2600) 

4150 
(3650) 

3800 
(3400) 

0.1 NPLC 

Internal 

580 
(480) 

560 
(470) 

550 
(465) 

550 
(460) 

560 
(470) 

545 
(460) 

0.1 NPLC 

Digital I/O 

550 
(460) 

550 
(460) 

540 
(450) 

540 
(450) 

560 
(470) 

545 
(460) 

1.0 NPLC 

Internal 

59 
(49) 

59 
(49) 

59 
(49) 

59 
(49) 

59 
(49) 

59 
(49) 

1.0 NPLC 

Digital I/O 

59 
(49) 

59 
(49) 

59 
(49) 

59 
(49) 

59 
(49) 

59 
(49) 

 
 

 

Maximum single measurement rates (operations per second) for 60 Hz (50 Hz): 

A/D converter speed 

Trigger origin 

Measure to GPIB 

Source measure to 
GPIB 

Source measure 
pass/fail to GPIB 

0.001 NPLC 

Internal 

1900 (1800) 

1400 (1400) 

1400 (1400) 

0.01 NPLC 

Internal 

1450 (1400) 

1200 (1200) 

1100 (1100) 

0.1 NPLC 

Internal 

450 (390) 

425 (370) 

425 (375) 

1.0 NPLC 

Internal 

58 (48) 

57 (48) 

57 (48) 

 

                                              

20

 

Tests performed with a Model 2602B using the following equipment: Computer hardware (Intel

®

 Pentium

®

 4 2.4 GHz, 2 GB RAM, National 

Instruments™ PCI-GPIB); driver (NI-488.2 Version 2.2 PCI-GPIB); software (Microsoft

®

 Windows

®

 XP, Microsoft

®

 Visual Studio

®

 2010, VISA™ 

version 4.1).

 

21

 

Exclude current measurement ranges less than 1 mA. 

Summary of Contents for System SourceMeter 2601B

Page 1: ...and measurement accuracies are specified at the Models 2601B 2602B and 2604B terminals under these conditions 1 23 C 5 C 70 percent relative humidity 2 After a two hour warm up period 3 Speed normal...

Page 2: ...pA 0 03 5 nA 60 pA 10 pA 0 025 1 5 nA 100 A 2 nA 0 03 60 nA 3 nA 100 pA 0 02 25 nA 1 mA 20 nA 0 03 300 nA 6 nA 1 nA 0 02 200 nA 10 mA 200 nA 0 03 6 A 200 nA 10 nA 0 02 2 5 A 100 mA 2 A 0 03 30 A 600...

Page 3: ...ibe the typical performance of the Models 2601B 2602B and 2604B PULSE CHARACTERISTICS Pulse region specifications Pulse region specifications Region quadrant diagram Region maximums Maximum pulse widt...

Page 4: ...HARACTERISTICS Noise 10 Hz to 20 MHz 20 mV peak peak 3 mV RMS 6 V range Transient response time 70 s for the output to recover to within 0 1 for a 10 to 90 step change in load Overshoot Voltage 0 1 10...

Page 5: ...nal value after source level command is processed on a fixed range Voltage range Settling time 100 mV 50 s 1 V 50 s 10 V 110 s 40 V 12 150 s Current source output settling time Time required to reach...

Page 6: ...quired to reach within 0 1 of final value after source level command is processed on a fixed range Values below for Vout 1 V Current range Settling time 1 mA 100 s Input impedance 10 G ADDITIONAL CHAR...

Page 7: ...ge Settling time 3 A and 1A 120 s Rload 2 100 mA and 10 mA 100 s 1 mA 3 ms 100 A 3 ms 10 A 230 ms 1 A 230 ms Capacitor leakage performance Using HIGH C scripts 19 200 ms 50 nA Load 5 F in parallel wit...

Page 8: ...01 NPLC Digital I O 3500 3100 3400 3000 3000 2700 2900 2600 4150 3650 3800 3400 0 1 NPLC Internal 580 480 560 470 550 465 550 460 560 470 545 460 0 1 NPLC Digital I O 550 460 550 460 540 450 540 450 5...

Page 9: ...maximum rate is 190 second Maximum source function change rate 1000 per second Command processing time 1 ms Maximum time required for the output to begin to change after receiving the smua source leve...

Page 10: ...ts for example branching looping and math Able to execute high speed test scripts stored in memory without host intervention Minimum user memory available 16 MB approximately 250 000 lines of TSP code...

Page 11: ...ink expansion interface allows TSP enabled instruments to trigger and communicate with each other See the figure below Each Model 2601B and 2602B has two TSP Link connectors to make it easier to conne...

Page 12: ...or higher LAN crossover cable 9 84 ft 3 m maximum between each TSP enabled instrument USB Control Rear USB 2 0 Device USB TMC488 protocol USB File System Front USB 2 0 Host Mass storage class device...

Page 13: ...current flowing out of digital I O bit 960 A Maximum sink current at maximum logic low voltage 0 7 5 0 mA Absolute maximum sink current flowing into digital I O pin 11 mA 5 V power supply pins Limited...