7-14
Status Structure
Models 2510 and 2510-AT User’s Manual
Measurement event register
The used bits of the Measurement Event Register (shown in
follows:
•
Bit B0, Over Temperature (OT)
— Set bit indicates an over temperature
condition.
•
Bit B1, Under Temperature (UT)
— Set bit indicates an under temperature
condition.
•
Bit B2, Over Voltage (OV)
— Set bit indicates an over voltage condition.
•
Bit B3, Current Overload (CO)
— Set bit indicates a current overload condition.
•
Bit B4, Over Resistance (OR)
— Set bit indicates an over resistance condition.
•
Bit B5, Under Resistance (UR)
— Set bit indicates an under resistance condition.
•
Bit B6, Reading Available (RAV)
— Set bit indicates that a reading was taken and
processed.
•
Bit B7, Reading Overflow (ROF)
— Set bit indicates that the reading exceeds the
selected measurement range of the Model 2510.
•
Bits B8 through B10
— Not used.
•
Bit B11, Output Enable (OE)
— Set bit indicates that the Digital I/O port output
enable line is asserted.
•
Bit B12, Setpoint Tolerance (SPT)
— Set bit indicates that the setpoint is within
the programmed tolerance for the required number of readings.
•
Bit B13, Open Lead (OL)
— Set bit indicates that the temperature sensor leads are
open.
•
Bit B14, Shorted Lead (SL)
— Set bit indicates that the temperature sensor leads
are shorted.
•
Bit B15
— Always zero.