Models 2510 and 2510-AT
Specifications
A-3
Accuracy calculations
The information below discusses how to calculate accuracy for both TEC measurement and
thermal feedback element specifications.
Accuracy specifications are stated as follows:
Accuracy = ±(% of r offset)
As an example of how to calculate the actual limits, assume an output voltage of 5V. You can
compute the limits from one-year operating voltage accuracy specifications as follows:
Accuracy =
±(% of r offset)
±[(0.1% X 5V) + 4mV]
±(5mV + 4mV)
±9mV
Thus, the actual operating voltage range is: 5V± 9mV, or from 4.991 to 5.009V.