Models 2510 and 2510-AT User’s Manual
Basic Operation
3-7
Table 3-4
Resistance configuration menu
Configuration menu
item
Description
FUNCTION
Select DC resistance function.
AC-OHMS
Select AC resistance function.
TEMP-CONTROL
PROTECTION
CONTROL
LOLIM
HILIM
Program DC resistance parameters.
Configure resistance protection.
ENABLE or DISABLE resistance protection.
Program lower resistance limit:
1
Ω
to 200
Ω
(100
Ω
RTD or thermistor)
5
Ω
to 2k
Ω
(1k
Ω
RTD or thermistor)
50
Ω
to 20k
Ω
(10k
Ω
thermistor)
500
Ω
to 200k
Ω
(100k
Ω
thermistor)
Program higher resistance limit:
50
Ω
to 1k
Ω
(100
Ω
thermistor)
500
Ω
to 9.999k
Ω
(1k
Ω
thermistor)
5k
Ω
to 80k
Ω
(10k
Ω
thermistor)
50k
Ω
to 200k
Ω
(100k
Ω
thermistor)
50
Ω
to 250
Ω
(100
Ω
RTD)
500
Ω
to 3k
Ω
(1k
Ω
RTD)
SENSOR-TYPE
THERMISTOR
RANGE
I-SRC
SENSE-MODE
RTD
I-SRC
RANGE
SENSE-MODE
Select sensor type and settings.
Program thermistor settings.
Set thermistor range (100
Ω
, 1k
Ω
, 10k
Ω
, 100k
Ω
).
Set thermistor current (AUTO, 3.3µA, 10µA, 33.3µA, 100µA, 833.3µA,
2.5mA).
Select 2-WIRE or 4-WIRE sense mode.
Program RTD settings.
Set RTD current (AUTO, 3.3µA, 10µA, 33.3µA, 100µA, 833.3µA, 2.5mA).
Set RTD range (100
Ω
or 1000
Ω
).
Select 2-WIRE or 4-WIRE sense mode.
PID
PROPORTIONAL
INTEGRAL
DERIVATIVE
Program PID loop constants for resistance.
Set proportional constant (0 to 100000).
Set integral constant (0 to 100000).
Set derivative constant (0 to 100000).
Summary of Contents for TEC SourceMeter 2510
Page 194: ...A Specifications...
Page 197: ...B StatusandErrorMessages...
Page 206: ...C IEEE 488BusOverview...
Page 221: ...D IEEE 488andSCPI ConformanceInformation...
Page 225: ...E ExamplePrograms...
Page 233: ...F GPIB488 1Protocol...