Step 8: Turn output off.
When finished, turn the output off and halt PID control by sending:
:OUTP OFF
Protection limits
Protection limits for the four functions are discussed below.
summarizes protec-
tion limit characteristics.
NOTE
For temperature and resistance, the upper limit cannot be set below the lower
limit and the lower limit cannot be set above the upper limit. Also, the setpoint
cannot exceed limit setting(s).
Table 3-8
Protection limit characteristics
Function
Limit range
Output
Inhibited?*
Display message
Temperature**
Lower limit: -50 to +250°C
Upper limit: -50 to +250°C
Yes
Yes
UNDER-TEMP
OVER-TEMP
Voltage
0.5 to 10.5V
No
VLIM
Current
1.0 to 5.25A
No
ILIM
Resistance**
Lower limit:
1
Ω
to 200
Ω
(100
Ω
RTD or thermistor)
5
Ω
to 2k
Ω
(1k
Ω
RTD or thermistor)
50
Ω
to 20k
Ω
(10k
Ω
thermistor)
500
Ω
to 200k
Ω
(100k
Ω
thermistor)
Upper limit:
50
Ω
to 1k
Ω
(100
Ω
thermistor)
500
Ω
to 9.999k
Ω
(1k
Ω
thermistor)
5k
Ω
to 80k
Ω
(10k
Ω
thermistor)
50k
Ω
to 200k
Ω
(100k
Ω
thermistor)
50
Ω
to 250
Ω
(100
Ω
RTD)
500
Ω
to 3k
Ω
(1k
Ω
RTD)
Yes
Yes
UNDER-OHMS
OVER-OHMS
* Yes indicates that output turns off if limit is exceeded and cannot be turned on when in limit.
**Lower limit cannot be set above upper limit. Upper limit cannot be set below lower limit.
Summary of Contents for TEC SourceMeter 2510