2-22
Calibration
Models 2510 and 2510-AT
Temperature
Thermistor
sensor
RTD sensor
Current sensor
Voltage sensor
:SYST:RSEN ON
:SENS:TEMP:TRAN THER
:SENS:TEMP:THER:RANG 100
:CAL:PROT:TEMP <Resistance_Value>
:CAL:PROT:TEMP 0
:SENS:TEMP:THER:RANG 1e3
:CAL:PROT:TEMP <Resistance_Value>
:CAL:PROT:TEMP 0
:SENS:TEMP:THER:RANG 1e4
:CAL:PROT:TEMP <Resistance_Value>
:CAL:PROT:TEMP 0
:SENS:TEMP:THER:RANG 1e5
:CAL:PROT:TEMP <Resistance_Value>
:CAL:PROT:TEMP 0
:SENS:TEMP:TRAN RTD
:SENS:TEMP:RTD:RANG 100
:CAL:PROT:TEMP <Resistance_Value>
:CAL:PROT:TEMP 0
:SENS:TEMP:RTD:RANG 1000
:CAL:PROT:TEMP<Resistance_Value>
:CAL:PROT:TEMP 0
:SENS:TEMP:TRAN ISS
:CAL:PROT:TEMP <Resistance_Value>
:CAL:PROT:TEMP <DMM_Reading>
:SENS:TEMP:TRAN VSS
:CAL:PROT:TEMP <DMM_Reading>
:CAL:PROT:TEMP 0
100
Ω
Short to INPUT (Figure 2-4)
1k
Ω
to INPUT (Figure 2-3)
Short to INPUT (Figure 2-4)
10k
Ω
Short to INPUT (Figure 2-4)
100k
Ω
Short to INPUT (Figure 2-4)
100
Ω
Short to INPUT (Figure 2-4)
1000¾ to INPUT (Figure 2-3)
Short to INPUT (Figure 2-4)
1k
Ω
/DMM to INPUT (Figure 2-5)
1k
Ω
/DMM to INPUT (Figure 2-5)
Short to INPUT (Figure 2-4)
AC ohms
:CAL:PROT:RES <Resistance_Value>
100
Ω
Voltage source
:OUTP ON
:CAL:PROT:SOUR:AUTO
:OUTP OFF
None (open OUTPUT)
Table 2-6 (cont.)
Remote calibration step summary
Calibrated function
Calibration command
Test connections
2510-902-01 (E - Mar 2010)(Service).book Page 22 Tuesday, March 16, 2010 1:04 PM
Summary of Contents for TEC SourceMeter 2510
Page 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Page 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Page 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...