Models 2510 and 2510-AT
Routine Maintenance
3-3
Perform the following steps to replace the line fuse:
1.
Using a small flat-blade screwdriver, carefully release the locking tab that secures the
fuse carrier to the power module.
2.
Pull out the fuse carrier, and replace the fuse with the type specified in Table 3-1.
CAUTION
To prevent instrument damage, use only the fuse type specified in
Table 3-1.
3.
Reinstall the fuse carrier, pushing it in firmly until it locks into place.
NOTE
If the power line fuse continues to blow, a circuit malfunction exists and must be
corrected. Refer to the troubleshooting information in Section 4 of this manual for
additional information.
Table 3-1
Power line fuse
Line voltage
Rating
Keithley part no.
100-240V
250V, 2.5A, Slow
Blow 5
×
20mm
FU-106-2.5
2510-902-01 (E - Mar 2010)(Service).book Page 3 Tuesday, March 16, 2010 1:04 PM
Summary of Contents for TEC SourceMeter 2510
Page 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Page 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Page 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...