1-7
Performance Verification
Models 2510 and 2510-AT
5.
Using the EDIT keys, adjust the Model 2510 voltage setpoint to e9.000V.
6.
Verify that the DMM reading is within 8.987 to 9.013V limits.
7.
Repeat steps 5 and 6 for a -9.000V output value.
8.
Turn off the output when the test is completed.
Voltage readback accuracy
Follow the steps below to verify that Model 2510 voltage readback accuracy is within spec-
ified limits. This test involves setting the output voltage to a specific value as measured by a
DMM and verifying that the Model 2510 voltage readback display reading is within limits.
1.
With the power off, connect the digital multimeter to the Model 2510 OUTPUT termi-
nals, as shown in Figure 1-1.
2.
Turn on the Model 2510 and DMM, and allow them to warm up for at least one hour.
3.
Select the multimeter DC volts measuring function, and choose the 20V range.
4.
Press the Model 2510 V key to select the voltage function, and make sure the source
output is turned on.
5.
Using the EDIT keys, adjust the Model 2510 output voltage as measured by the DMM
to +9.000V (or closest possible value).
6.
Verify that the Model 2510 readback display (top display line) is within 8.987 to
9.013V limits.
7.
Repeat steps 5 and 6 for a -9.000V output value.
8.
Turn off the output when the test is completed.
Figure 1-1
Connections for voltage
verification tests
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
120
Model 2510
OUTPUT F-
LINE RATING
100-240VAC
50, 60 HZ
90VA MAX
LINE FUSE
SLOWBLOW
2.5A, 250V
!
!
CAT
I
OUTPUT S-
OUTPUT F+
OUTPUT S+
NEXT
PREV
POWER
DISPLAY
2002
MULTIMETER
DCV
ACV
DCI
ACI
Ω
2
Ω
4
FREQ
TEMP
REL
TRIG
INFO
LOCAL
EXIT
ENTER
RANGE
RANGE
!
F
R
500V
PEAK
FRONT/REAR
2A 250V
AMPS
CAL
STORE
RECALL
CHAN
SCAN
FILTER
MATH
CONFIG
MENU
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1100V
PEAK
AUTO
Model 2002 DMM
INPUT HI
INPUT LO
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
MADE IN
U.S.A.
ENABLE-DIG I/O
RS-232
TRIGGER
LINK
OUTPUT
F+ S+ S- F-
F+ F- S+ S-
ISOLATION FROM EARTH: 30V MAX.
INPUT
2510-902-01 (E - Mar 2010)(Service).book Page 7 Tuesday, March 16, 2010 1:04 PM
Summary of Contents for TEC SourceMeter 2510
Page 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Page 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Page 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...