Models 2510 and 2510-AT
Calibration
2-13
NOTE
The characterized 1
Ω
resistor can be used in place of the shorted INPUT terminals,
if desired. Enter the characterized value of the 1
Ω
resistor at the RES VALUE
prompt.
8.
Press ENTER. The Model 2510 will display the following:
RES VALUE: 00.00000
Ω
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
9.
Press ENTER to complete the shorted calibration step, then remove the shorting wires
from the INPUT terminals.
10.
From the THERMISTOR CAL menu, select 1k
Ω
, then press ENTER. The unit will dis-
play the following:
THERMISTOR CAL
Connect 1k
Ω
to temperature
©
ß
sensor and press ENTER.
11.
Connect the 1k
Ω
resistor to the Model 2510 INPUT terminals (see Figure 2-3).
12.
Press ENTER. The Model 2510 will display the following:
RES VALUE: 01.00000k
Ω
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
13.
Use the EDIT keys to adjust the Model 2510 display to agree exactly with the charac-
terized 1k
Ω
resistance value, then press ENTER. The unit will display the following:
THERMISTOR CAL
Connect a Short to temperature
©
ß
sensor and press ENTER.
14.
Short all four INPUT terminals together using clean copper wire (see Figure 2-4).
Allow one minute for thermal equilibrium.
15.
Press ENTER. The Model 2510 will display the following:
RES VALUE: 00.00000
Ω
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
16.
Press ENTER to complete the shorted calibration step, then remove the shorting wires
from the INPUT terminals.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
120
LINE RATING
100-240VAC
50, 60 HZ
90VA MAX
LINE FUSE
SLOWBLOW
2.5A, 250V
!
!
CAT
I
Model 2510
Short INPUT Terminals
with Clean Copper Wire
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
MADE IN
U.S.A.
ENABLE-DIG I/O
RS-232
TRIGGER
LINK
OUTPUT
INPUT
F+ S+ S- F- F+ F- S+ S-
ISOLATION FROM EARTH: 30V MAX.
Figure 2-4
Shorted input
calibration connections
2510-902-01 (E - Mar 2010)(Service).book Page 13 Tuesday, March 16, 2010 1:04 PM
Summary of Contents for TEC SourceMeter 2510
Page 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Page 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Page 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...