2-18
Calibration
Models 2510 and 2510-AT
Step 5: AC ohms calibration
1.
From the CAL EXECUTE menu, select AC-OHMS, then press ENTER. The instru-
ment will display the following message:
AC-OHMS CAL
Connect a 100
Ω
load to Peltier.
©
ß
Then press ENTER.
2.
Connect the 100
Ω
resistor to the OUTPUT terminals (see Figure 2-6).
3.
Press ENTER. The instrument will display the following:
RES VALUE: 0.100000k
Ω
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
4.
Use the EDIT keys to adjust the Model 2510 display to agree exactly with the charac-
terized 100
Ω
resistance value, then press ENTER.
Step 6: Voltage source calibration
1.
From the CAL EXECUTE menu, select V-SOURCE, then press ENTER. The instru-
ment will display the following message:
V-SRC CAL
Open Peltier leads then press
©
ß
ENTER.
2.
Disconnect all resistors and test leads from the Model 2510 OUTPUT terminals.
3.
Press ENTER to complete voltage source calibration.
4.
Press EXIT to return to the CALIBRATION menu.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
120
LINE RATING
100-240VAC
50, 60 HZ
90VA MAX
LINE FUSE
SLOWBLOW
2.5A, 250V
!
!
CAT
I
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
MADE IN
U.S.A.
ENABLE-DIG I/O
RS-232
TRIGGER
LINK
OUTPUT
INPUT
F+ S+ S- F- F+ F- S+ S-
ISOLATION FROM EARTH: 30V MAX.
Model 2510
100
Ω
Resistor
OUTPUT S-
OUTPUT F-
OUTPUT S+
OUTPUT F+
Connect sense (S) leads as close
as possible to resistor body.
Figure 2-6
AC ohms calibration
connections
2510-902-01 (E - Mar 2010)(Service).book Page 18 Tuesday, March 16, 2010 1:04 PM
Summary of Contents for TEC SourceMeter 2510
Page 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Page 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Page 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...