5-4
Disassembly
Models 2510 and 2510-AT
Case cover removal
Follow the steps below to remove the case cover to gain access to internal parts.
WARNING
Before removing the case cover, disconnect the line cord and any test leads
from the instrument.
1.
Remove handle
— The handle serves as an adjustable tilt-bail. Adjust its position by
gently pulling it away from the sides of the instrument case and swinging it up or down.
To remove the handle, swing the handle below the bottom surface of the case and back
until the orientation arrows on the handles line up with the orientation arrows on the
mounting ears. With the arrows lined up, pull the ends of the handle away from the
case.
2.
Remove mounting ears
— Remove the screw that secures each mounting ear. Pull
down and out on each mounting ear.
NOTE
When reinstalling the mounting ears, make sure to mount the right ear to the right
side of the chassis, and the left ear to the left side of the chassis. Each ear is marked
“RIGHT” or “LEFT” on its inside surface.
3.
Remove rear bezel
— To remove the rear bezel, loosen the two screws that secure the
rear bezel to the chassis, then pull the bezel away from the case.
4.
Remove bottom screws
— Remove the four screws that secure the case to the chassis.
They are located on the bottom of the case.
5.
Remove chassis
— To remove the case, grasp the front bezel of the instrument, and
carefully slide the chassis forward. Slide the chassis out of the metal case.
Mother board removal
Perform the following steps to remove the mother board. This procedure assumes that the
case cover is already removed.
1.
Remove the IEEE-488, ENABLE-DIG I/O, and RS-232 fasteners.
The IEEE-488, ENABLE-DIG I/O, and RS-232 connectors each have two hex-head
screws that secure the connectors to the rear panel. Remove these screws.
2.
Remove mother board mounting screws.
Remove the two mounting screws that secure the mother board to the chassis.
3.
Unplug cables:
•
Unplug the display board ribbon cable from J1014.
•
Unplug the cables going to the power supply from J1003.
•
Unplug the cable going to the OUTPUT indicator from J102.
•
Unplug the fan cable from J1015.
2510-902-01 (E - Mar 2010)(Service).book Page 4 Tuesday, March 16, 2010 1:04 PM
Summary of Contents for TEC SourceMeter 2510
Page 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Page 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Page 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Page 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...