Models 2510 and 2510-AT User’s Manual
Getting Started
1-17
Table 1-3
Factory front panel default settings
Setting
Front panel BENCH and GPIB default*
Control function
Temperature
Temperature function
Setpoint
25°C (298.15K, 77°F)
Protection high limit
50°C (323.15K, 122°F)
Protection low limit
0°C (273.15K, 32°F)
Protection state
Enabled
Temperature units
Celsius
Sensor current
2.5mA (100
Ω
range, RTD)
833.3µA (1k
Ω
range)
100µA (10k
Ω
range)
33.3µA (100k
Ω
range)
Sensor auto current state
On
Sensor type
Thermistor
Sense mode
2-wire
RTD sensor parameters
Type
PT385
Range
100
Ω
Alpha (
α
)
0.00385
Beta (
β
)
0.111
Delta (
δ
)
1.507
Thermistor sensor parameters
Range
10k
Ω
A
1.13030e-3
B
2.33894e-4
C
8.85983e-8
Solid-state sensor parameters
VSS gain
1e-2
VSS offset
0
ISS gain
1e-6
ISS offset
0
Proportional constant
20
Integral constant
0.6
Derivative constant
0
Voltage function
Setpoint
0V
Voltage protection limit
10.5V
Proportional constant
10
Integral constant
20
Derivative constant
0
*
BENCH defaults are the same as :SYSTem:PRESet defaults (
). GPIB defaults are the same
as *RST defaults (
** Not affected by BENCH or GPIB RESET.