9. OPTIONS
3) Adjusting the RF frequency
OSet the PATTERN switch to "COLOR" and switch the
LUMI OFF and CHROMA OFF switches to "ON".
OSet the RF switch on the front panel to "ON" and
connect a counter to the output terminal. Adjust
VR15 for CH A and VR16 for CH B. It the RF level is
lower than the counter trigger level, introduce a high-
frequency amp into the connection with the counter.
OBe sure to use a terminal resistance of 75 ohms.
RT-62 Remote Controller
A hand-held controller that allows remote operation of
the instrumentation. The controller contains two con-
trol systems which use 4-bit binary signals, and it
allows switching to be accomplished among 16 ranges
and modes in as many as eight steps each. In addition,
one of the control systems can be programmed to
accommodate a 16-step mode, which can be selected
by a push- switch.