BHK-MG (OPR) 032816
3-69/(3-70 Blank)
3. Program slave unit as follows:
a. Program output current of slave unit (Iset
S
) to 0.
b. Program Vset
S
to the rated maximum voltage of the slave unit, Eomax
S
(e.g., for
BHK1000-0.2MG set Vset
S
= 1000).
c. Program overcurrent value of slave unit (Iocset
S
) to 1.1 x rated maximum current of
slave unit, Iomax
S
(e.g., for BHK 1000-0.2MG, program Iocset
S
= 220 (mA)).
d. Program overvoltage value of slave unit (Vovset
S
) to 1.1 x voltage limit E
LIM
desired for
power assembly.
4. Program master unit as follows:
a. Program Iset
M
to 1/2 the desired output current, Io, of the power assembly.
b. Program Vset
M
to the voltage limit, E
LIM
desired for power assembly.
c. Program overcurrent value of master unit (Iocset
M
) to 1.1 x Iset
M
(step 4a).
d. Program overvoltage value of master unit (Vovset
M
) to 1.1 x Vset
M
(step 4b).
5. To apply power to the load:
a. If load switch used, open the switch.
b. If load switch not used, press POWER ON/OFF key once on slave unit, then once on
master unit.
6. To disable power to the load:
a. If load switch used, close the switch.
b. If load switch not used, press POWER ON/OFF key once on master unit, then once on
slave unit.
c. Alternative: Turn both units off by setting POWER switch to off (down) position, first on
master unit, then on slave unit.
Summary of Contents for BHK 1000-0.2MG
Page 2: ......
Page 6: ......
Page 17: ...BHK MG032816 xi FIGURE 1 1 BHK MG SERIES PROGRAMMABLE POWER SUPPLY ...
Page 18: ...xii BHK MG032816 ...
Page 46: ......
Page 101: ...BHK MG OPR 032816 3 55 FIGURE 3 20 SERIES AUTOMATIC CONFIGURATION ...
Page 109: ...BHK MG OPR 032816 3 63 FIGURE 3 23 PARALLEL AUTOMATIC CONFIGURATION ...
Page 114: ...3 68 BHK MG OPR 032816 FIGURE 3 25 PARALLEL MASTER SLAVE CURRENT MODE CONFIGURATION ...
Page 116: ......
Page 154: ......