BHK-MG (OPR) (OPR) 032816
4-7
the output by a maximum of 18 steps (each step is one LSB (Least Significant Bit), equivalent to
0.024% of nominal value) at a time; the
>
and
<
buttons either increase (
>
) or decrease (
<
) the
output one step at a time.
FIGURE 4-3. CALIBRATION CONTROLS
4.4.4
CALIBRATION PROCEDURE
1. Click on the
Calibrate
button on the main control panel (see Figure 4-2). This opens a cali-
bration panel (see Figure 4-3 and Table 4-2) that allows calibration in either voltage or cur-
rent mode.
2. After initializing, the password window appears (see Figure 4-4). The password has been set
at the factory (see Table 4-3). Type the proper password in the text box and click
OK
. If the
password has been forgotten, consult factory. If the password is correct, calibration can be
accomplished by following the screen prompts.
TABLE 4-2. CALIBRATION PANEL FUNCTIONS
BUTTON OR WINDOW
Function
<<
button
Coarse Adjustment - Decreases the output value by a maximum of 18 LSBs.
>>
button
Coarse Adjustment - Increases the output value by a maximum of 18 LSBs.
<
button
Fine Adjustment - Decreases the output value by one LSB.
>
button
Fine Adjustment - Increases the output value by one LSB.
OK button
Enters the value displayed in the text field and advances calibration to next step.
Summary of Contents for BHK 1000-0.2MG
Page 2: ......
Page 6: ......
Page 17: ...BHK MG032816 xi FIGURE 1 1 BHK MG SERIES PROGRAMMABLE POWER SUPPLY ...
Page 18: ...xii BHK MG032816 ...
Page 46: ......
Page 101: ...BHK MG OPR 032816 3 55 FIGURE 3 20 SERIES AUTOMATIC CONFIGURATION ...
Page 109: ...BHK MG OPR 032816 3 63 FIGURE 3 23 PARALLEL AUTOMATIC CONFIGURATION ...
Page 114: ...3 68 BHK MG OPR 032816 FIGURE 3 25 PARALLEL MASTER SLAVE CURRENT MODE CONFIGURATION ...
Page 116: ......
Page 154: ......