BHK-MG (OPR) (OPR) 032816
4-11
4.6
RESTORING PREVIOUS CALIBRATION VALUES
Each time the unit is calibrated, the previous calibration values are retained in non-volatile mem-
ory. If it is determined that a new calibration is erroneous, the previous calibration values can be
restored.
1. With the power supply in command entry status
(:_:_:)
, press
MENU
key until the LCD
reads
GoTo Previous CALIB? 1(=_=_=)YES
.
2. Press
1
, then ENTER. LCD reads
Calibration ! PassWord(=_=_=)
.
3. Enter 4-digit password and press
ENTER
. Previous calibration values are restored.
4.7
RESTORING FACTORY CALIBRATION VALUES
The original factory calibration values can be restored.
1. With the power supply in command entry status
(:_:_:)
, press
MENU
key until the LCD
reads
GoTo Factory CALIB? 1(=_=_=)YES
.
2. Press 1, then
ENTER
. LCD reads
Calibration ! PassWord(=_=_=)
.
3. Enter 4-digit password and press
ENTER
. Factory calibration values are restored.
4.8
CALIBRATION STORAGE
The BHK-MG Power Supply. maintains the calibration tables in Flash Memory until a PACK is
executed. There are six calibration areas maintained in Flash Memory: Working, Prior, Oldest,
Factory, Master, and First. The calibration password is required In order to store the calibration.
The calibration can be copied to another area using the CAL:DUMP? and CAL:COPY com-
mand. The syntax is as follows: CAL:DUMP? [source];:CAL:COPY [destination], where [source]
and [destination] refer to the areas of Flash memory where calibration data is stored, designated
as: WORKing, PRIor, OLDest, FACTory, MASTer, FIRst. Source refers to the calibration area
that is to be copied, destination to the area that the calibration will be copied into.
The ability to save to the FACTORY area is prohibited. If a user needs to overwrite the factory
calibration, a special firmware needs to be created to provide a special password for this capa-
bility. Each customer that requests this capability will be provided a password that the company
can specify. it must contain at least one alpha, one numeric and one punctuation character
(!@#$%^&*) and be at least 6 characters long and not longer than 20 characters in length. All
alpha data must be upper case and no control nor space characters are allowed.
The Master calibration should never be overwritten. Factory, Master, and First are set to the
same values when the power supply is calibrated at the factory. The Working calibration is the
active calibration. Each time a CAL:SAV is executed, the values are saved in the Working
(active) area. At the same time, the values previously stored in Working are moved to Prior, and
the values previously stored in Prior are moved to Oldest. The values previously stored in Old-
est are no longer available. Table 4-4 illustrates calibration storage and the use of the
CAL:DUMP? query and CAL:COPY command.
Summary of Contents for BHK 1000-0.2MG
Page 2: ......
Page 6: ......
Page 17: ...BHK MG032816 xi FIGURE 1 1 BHK MG SERIES PROGRAMMABLE POWER SUPPLY ...
Page 18: ...xii BHK MG032816 ...
Page 46: ......
Page 101: ...BHK MG OPR 032816 3 55 FIGURE 3 20 SERIES AUTOMATIC CONFIGURATION ...
Page 109: ...BHK MG OPR 032816 3 63 FIGURE 3 23 PARALLEL AUTOMATIC CONFIGURATION ...
Page 114: ...3 68 BHK MG OPR 032816 FIGURE 3 25 PARALLEL MASTER SLAVE CURRENT MODE CONFIGURATION ...
Page 116: ......
Page 154: ......