B-12
BHK-MG (OPR) 032816
B.49
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPlitude]? QUERY
CURR:TRIG?
Syntax:
Short Form: [SOUR:]CURR[:LEV]:TRIG[:AMP]?
Long Form: [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPlitude]?
Return Value: <exp_value> = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1
Description:
Returns value representing voltage value to be programmed by *TRG command established by
CURR:TRIG command).
(See example, Figure B-1.)
FIGURE B-4. PROGRAMMING CURRENT
B.50
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude] COMMAND
VOLT
Syntax:
Short Form: [SOUR:]VOLT[:LEV][:IMM][:AMP] <exp_value>
Long Form: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude] <exp_value>
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1
Description:
Sets programmed voltage level at power supply output.
This command programs output voltage
to a specific value; actual output voltage will depend on load conditions. If the value exceeds the
VOLT:LIM[:HIGH] value, error message -222,”Data out of range” is posted in output queue. Related
Commands: *RCL, *SAV, VOLT:LIM[:HIGH]. (See example, Figure B-1.
B.51
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude]? QUERY
VOLT?
Syntax:
Short Form: [SOUR:]VOLT[:LEV][:IMM][:AMP]? {MIN | MAX}
Long Form: [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPlitude]? {MIN | MAX}
Description:
Identifies programmed voltage, maximum allowable voltage, or minimum voltage (always 0).
The VOLT? query returns the programmed value of voltage. Actual output voltage will depend on load
conditions. The VOLT?MAX query returns the maximum voltage allowed for a particular model (e.g.,
500V for BHK 500-80MG). VOLT? MIN returns minimum voltage allowed for power supply (always 0).
Related Commands: VOLT. (See example, Figure B-5.)
B.52
[SOURce:]VOLTage:LIMit[:HIGH] COMMAND
VOLT:LIM[:HIGH]
Syntax:
Short Form: [SOUR:]VOLT:LIM[:HIGH] <exp_value>
Long Form: [SOURce:]VOLTage:LIMit[:HIGH] <exp_value>
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1
Description:
Sets voltage limit for power supply output.
If the value exceeds the voltage protection value (set by
VOLT:PROT, see PAR. B.56), error message -222,”Data out of range” is posted in the output queue.
NOTE: If the voltage limit is changed, any previously established trigger levels higher than the new
limit must be reprogrammed to be lower than the new limit; otherwise triggers will cause the output to
exceed the new limit. Once the limit is established, the unit will not accept trigger values higher than
the programmed limit. Related Commands: VOLT. (See example, Figure B-5. See PAR. 3.3.6.2 and
Figure 3-2 for special programming considerations.
NOTES: 1. The power supply is assumed to be operating in constant voltage (CV) mode.
2 Examples below are intended only to illustrate command functions. Refer to PAR. 3.3.6 for
programming techniques to optimize performance.
OUTP ON
Output enabled.
VOLT 421;CURR 1.1E-2
Power supply programmed to voltage limit 421V, 0.011A.
CURR?
Returns 1.1E-2.
CURR:LIM[:HIGH] 3.3E-2
Current limit set to 0.033A.
CURR:LIM[:HIGH]?
Returns 3.3e-2
CURR 4.2E-1
Power supply output current stays at 0.011A, error message
-222 posted.
CURR?
Returns 1.1e-2
CURR 1.0E-2
Power supply output current programmed to 0.01A
CURR?
Returns 1.0e-2 (0.01A).
CURR:PROT 1.25E-1
Unit shuts down if current exceeds 125 mA.
Summary of Contents for BHK 1000-0.2MG
Page 2: ......
Page 6: ......
Page 17: ...BHK MG032816 xi FIGURE 1 1 BHK MG SERIES PROGRAMMABLE POWER SUPPLY ...
Page 18: ...xii BHK MG032816 ...
Page 46: ......
Page 101: ...BHK MG OPR 032816 3 55 FIGURE 3 20 SERIES AUTOMATIC CONFIGURATION ...
Page 109: ...BHK MG OPR 032816 3 63 FIGURE 3 23 PARALLEL AUTOMATIC CONFIGURATION ...
Page 114: ...3 68 BHK MG OPR 032816 FIGURE 3 25 PARALLEL MASTER SLAVE CURRENT MODE CONFIGURATION ...
Page 116: ......
Page 154: ......