34
MBB-M-BA-e-1617
15 Instant help
In case of a fault in the program sequence, the balance should be shortly switched
off. The weighing process must then be restarted from the beginning.
Failure:
Possible cause:
The displayed weight
does not glow.
The balance is not switched on.
The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
Check fuse of adapter / glowing green LED next to
fuse
Power supply interrupted.
Batteries are inserted incorrectly or empty
No batteries inserted.
The displayed weight is
permanently changing
Draught/air movement
Table/floor vibrations
The seat surface/weighing plate is in contact with
foreign bodies or is not correctly positioned.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
The weighing result is
obviously incorrect
The display of the balance is not at zero.
Adjustment is no longer correct.
Great fluctuations in temperature.
The balance is on an uneven surface.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
Should other error messages occur, switch balance off and then on again. If the
error message remains inform manufacturer.
Summary of Contents for KERN MBB-M series
Page 41: ...MBB M BA d 1617 40 Die Auswahl erfolgt mit der HOLD und TARE Taste Werkseinstellung ...
Page 82: ...40 MBB M BA e 1617 Press the HOLD and TARE buttons to make the selection default setting ...
Page 85: ......
Page 107: ...MBB M BA f 1617 23 Branchement CN 3 de pile rechargeable Branchement CN 4 de piles AA x 6 ...
Page 125: ...MBB M BA f 1617 41 La sélection s opère sur la touche HOLD et TARE réglé en usine ...