background image

OBE-1-BA-e-1510

 

 

  10 

5.4  Adjust the brightness 

 

Adjust condenser aperture

 

Adjust the height of the condenser until the image from the field diaphragm appears 
clearly  in  the  visual  field.  For  some  microscopes  there  is  a  risk  that  you  will  lift  the 
condenser  up  so  high  that  it  collides  with  the  object  holder.  Care  is  needed  when 
doing this. 
 
For each slide,  use the 

aperture  diaphragm

 of the condenser to find the very best 

compromise  between contrast  and  brightness  for  the  microscopic  image.  The  scale 
on the condenser is to be regarded as a recommendation for each lens. 
 
It is possible to alter the brightness of the bulb using the 

dimmer

. The brightness is 

always controlled by the bulb brightness. 
 
It is possible to re-adjust the focus and x-y axis. 
 
Observe the object. 
 
 

5.5  Dioptre compensation and focussing 

 

For models with 

dioptre setting

Keep the eye on the fixed, non-adjustable lens barrel open and close the eye which 
is on the adjustable lens barrel. Then adjust the focus for the open eye using the fine 
adjustment knob. 
 
Then  open  the  other  eye  and  shut  the  eye  which  had  been  open.  Then  adjust  the 
focus  for  the  second  eye  on  the  dioptre  setting.  When  doing  this,  do  not  touch  the 
fine adjustment knob. 

 

If  necessary, 

alter  the  brake  action  for  the  control  knobs

  using  the  locking  ring. 

The  rotary  knobs  on  the  lefthand  side  are  connected  to  the  rotary  knobs  on  the 
righthand side. Please only rotate on one side at a time, do not rotate right and left at 
the same time in opposite directions, otherwise this will damage the drive. 
 
Turning  the lens revolving unit will bring the lens with the 

next magnification level

 

into  position.  Make  sure  that  the  lens  clicks  properly  into  place.  When  rotating  the 
lenses, check from the side to make sure that the glass holder does not get damaged 
by  the  lens.  Under  normal  conditions  you  should  be  able  to  select  the  next  lens 
immediately  and  the  focus  should  be  roughly  at  the  same  level.  Make  additional 
adjustments  with  the  fine  adjustment  knob.  If  when  you  rotate  the  lens  it  hits  the 
object  holder,  you  must  first  lower  the  specimen  stage  and  re-adjust  it.  For  lenses 
which work with 

oil immersion (these are marked with the word “oil”), first put the oil 

drops on the slide, bring the lens into position and move it closer to the object holder. 
The link between the object holder and lens is the oil.  After use or before changing 
the slide, remove the oil using a dust-free cloth. 

 

 

Summary of Contents for OBE 111

Page 1: ...mbH Ziegelei 1 D 72336 Balingen E mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com User instructions Compound laboratory microscope KERN OBE 111 OBE 112 O...

Page 2: ...al data model outfit 7 4 Assembly 8 4 1 Lenses 8 4 2 Condenser 8 4 3 Head 8 4 4 Eyepieces 8 5 Operation 9 5 1 Getting started 9 5 2 Adjust the gap between the eyes only for OBE 112 OBE 113 9 5 3 Adjus...

Page 3: ...itional risk of triggering an electric shock 1 2 Notes on the electrical system Before connecting to a mains power supply you must make sure that you are using the correct input voltage The informatio...

Page 4: ...fore you wipe down the device you must ensure that the mains power is switched off When glass components become dirty the best way to clean them is to wipe them gently with a lint free cloth To wipe o...

Page 5: ...sting screw Head Lens revolving unit Lens Object holder Specimen stage Aperture adjustment Condenser Condenser Field lens Control dial X Y axis Specimen stage Fine adjustment knob Coarse adjustment kn...

Page 6: ...OBE 1 BA e 1510 6 Rear view Mains connection Coarse adjustment knob Fine adjustment knob Fuse Condenser Focus dial Adjusting ring torque Example OBE 111 monocular...

Page 7: ...7 OBE 1 BA e 1510 3 Technical data model outfit...

Page 8: ...fitted in the device The height can be adjusted using the focus dial 4 3 Head The head is already fitted on the device After loosening the adjusting screw the head can be rotated by 180 degrees for ex...

Page 9: ...operly clicked into place Place the test object on the specimen stage and fix it with the object holder if necessary Use the control dial for the X Y axis specimen stage to position the sample in the...

Page 10: ...us for the second eye on the dioptre setting When doing this do not touch the fine adjustment knob If necessary alter the brake action for the control knobs using the locking ring The rotary knobs on...

Page 11: ...e microscope 7 Maintenance These microscopes are generally maintenance free Service work should be limited to the replacement of bulbs light sources and electrical fuses Should a defect occur please c...

Page 12: ...tly The condenser is too low The visual field is dark or is not correctly illuminated The lens has not been correctly adjusted The object revolving unit is not correctly fitted The condenser is not co...

Page 13: ...nt lens of the condenser Dirt dust on the object One side of the image is blurred The stage has not been assembled correctly The lens is not positioned correctly on the beam path The lens revolving un...

Page 14: ...materials which you can dispose of at your local recycling centre Disposal of the storage box and device must be carried out by the operator in accordance with all national or regional laws in force...

Reviews: