ODC-87_88-BA-e-1610
3
Scope of delivery
-
Microscope camera
-
USB cable
-
Object micrometre for calibration
-
Software CD
-
Adjustment rings (
Ø
30.0 mm +
Ø
30.5 mm) for eyepiece adapter
Mounting
1. Remove the black cover at the bottom of the camera.
2. The round connecting piece, where the cover was attached, has a
standardised diameter (Ø 23.2 mm). Thus the camera suites all
microscopes of which the eyepieces have this standard size.
3. For the mounting to the microscope one of the eyepieces needs to
be removed from the microscope tube and replaced by the
eyepiece camera.
Important:
For microscopes, that have a different eyepiece diameter (30.0
mm or 30.5 mm, mostly used for stereo microscopes), you need
to apply the appropriate adjustment ring in order to mount the
eyepiece camera correctly.
4. If necessary, adjust the microscope according to the trinocular
usage (with help of the trino toggle rod / trino toggle wheel).
PC connection
1. Establish USB connection via USB cable.
2. Installing the software with help of the CD.
Detailed instructions can be found in the supplied "readme"-file.
3. Both the supplied "Help"-files and the software-internal "User
Guide" include all information and instructions about the operation
of the software or of digital microscopy.